  • Triaxial ICP® miniature vibration sensor,
  • Measuring range 100 g,
  • Sensitivity 50 mV / g,.
  • Frequency range 1... 4,000 Hz,
  • 4-pin female connector on the side

 

Basic data
Sensitivity (±10 %) 50 mV / g 5.1 mV/(m/s²)  
 
Measuring range ±100 gPK ±981 m/s²PK  
 
Frequency range (≤5 %) 1... 4,000 Hz    
 
Frequency range (±10 %) 0.7... 5,000 Hz    
 
Resonance frequency ≥25. 000 Hz    
 
Wide band resolution 0.0003 g RMS 0.003 m/s² RMS [3]
 
Linearity error ±1 %   [1]
 
Transverse acceleration sensitivity ≤5 %    
Conditions of use
Overload limit (shock) ±5.000 gPK ±49.050 m/s²PK  
 
Temperature range (operation) -65... 250 ° F -54... 121 ° C
 
strain sensitivity (footprint) 0.001 g/µε 0.01 (m/s²)/µε [3]
specifications
power supply 23... 30 V DC   [2]
 
constant power supply 2... 20 mA    
 
output impedance ≤200Ω    
 
Bias voltage 7... 16 V DC    
 
discharging charging time constant 0,4 … 1,5 s    
 
settling time
(auf Bias-Spannung ±10 %)
  
 
spectral noise (1 Hz) 190 µg/√Hz 1,864 (µm/s²)/√Hz [3]
 
spectral noise (10 Hz) 35 µg/√Hz 345 (µm/s²)/√Hz [3]
 
spectral noise (100 Hz) 20 µg/√Hz 196 (µm/s²)/√Hz [3]
 
Spectral noise (1,000 Hz) 3 µg/√Hz 29.4 (µm/s²)/√Hz [3]
 
Spectral noise (10,000 Hz) 3 µg/√Hz 29.4 (µm/s²)/√Hz [3]
Physical properties
Sensor element Ceramic    
 
Geometry of the sensor element Shear principle    
 
Housing material Titanium    
 
Sealing Hermetic    
 
Size - height 0.45 in 11.4 mm  
 
Size - length 0.82 in 20.8 mm  
 
Size - width 0.45 in 11.4 mm  
 
Weight (without cable) 0.23 oz 6.6 grams [3]
 
Electrical connection 1/4 "-28 4-pin    
 
Connecting position On the side    
 
Fixing Adhesive mounting    
Notes
[1] Zero-balance just, determined with the method of klformer squares    
 
[2] The full range can be used only when a sufficiently high voltage.    
 
[3] Typically    
 
[4] Details: see PCBs Declaration of conformance PS023    
Included accessories
080A109 Mount wax -1  
 
ACS-1T Frequency response for Triaxial accelerometer, traceable to NIST, 10 Hz up to the upper 5%-Frequenz -1  
(HT) High temperature electronics
Sensitivity (±20 %) 50 mV / g 5.1 mV/(m/s²)  
 
Frequency range (≤5 %) 4... 4,000 Hz    
 
Frequency range (±10 %) 3... 5,000 Hz    
 
Temperature range 65... 325 ° F -54... 163 ° C  
 
Discharge charge time constant 0.1... 0.6 s    
(T) TEDS storage according to IEEE PP1451. 4
Supply voltage 24... 30 V DC    
 
Bias voltage 7.5... 17 V DC    
(TLA) TEDS LMS International - free format
Supply voltage 24... 30 V DC    
 
Bias voltage 7.5... 17 V DC    
(TLB) TEDS LMS international - format for the automotive industry
Supply voltage 24... 30 V DC    
 
Bias voltage 7.5... 17 V DC    
(TLC) TEDS LMS international - format for the aviation industry
Supply voltage 24... 30 V DC    
 
Bias voltage 7.5... 17 V DC    
(TLD) TEDS memory in accordance with IEEE P1451. 4
Supply voltage 24... 30 V DC    
 
Bias voltage 7.5... 17 V DC    
PCB-356A34