PCB-(M)603C11
zoom_out_map

PCB-(M)603C11

  • Industrial ICP® vibration sensor,
  • Measurement range 50 g,
  • Sensitivity 100 mV/g,
  • Frequency range 1.2 … 10,000 Hz,
  • One-point calibration at 100 Hz,
  • Integrated cable, length 3 m,
    ending on cut end
  •  

    basic data
    sensitivity (±10%) 100 mV/g 10.2 mV/(m/s²) [6]
     
    measuring range ±50 g ±490 m / s²  
     
    Frequency range (± 3 dB) 30 … 600,000 cpm 0.5 … 10,000 Hz [5]
     
    resonance frequency 1,500 kcpm 25,000 Hz [2]
     
    Wide band resolution 350 µg 3.434 µm / s ² [2]
     
    linearity ≤1%   [3]
     
    Transverse acceleration sensitivity ≤ 7%    
    conditions of use
    Overload limit (shock) 5,000 g pk 49,050 m/s² pk  
     
    temperature range -65 … 250 ° F -54 … 121 ° C  
     
    Type of protection IP68    
    Technical characteristics
    settling time (on bias voltage ± 1%) ≤2.0 s    
     
    discharging charging time constant ≥0.3 s    
     
    Power 18 … 28 V DC    
     
    constant power supply 2 … 20 mA    
     
    Output impedance < 150 Ω    
     
    Bias voltage 8 … 12 V DC    
     
    Spectral noise (10 Hz) 8 µg / √Hz 78.5 (µm / s ²) / √Hz [2]
     
    spectral noise (100 Hz) 5 µg / √Hz 49.1 (µm / s ²) / √Hz [2]
     
    Spectral noise (1,000 Hz) 4 µg / √Hz 39.2 (µm / s ²) / √Hz [2]
     
    electrical insulation (housing) >100 MΩ    
    Physical properties
    size - Allen 11/16 in 18 mm  
     
    size - height 4.7 in 119 mm  
     
    Dimensions (without cable) 1.8 oz 51 grams  
     
    mounting thread 1/4 "-28 threaded   [4]
     
    Assembly torque 2 … 5 ft-lb 2.7 … 6.8 Nm  
     
    sensor element ceramic    
     
    geometry of the Sensor element shear principle    
     
    housing material stainless steel    
     
    sealing hermetically sealed welded    
     
    Electrical connection potted integrated cable    
     
    cable outlet top    
     
    Cable length 10 ft 3.0 m  
     
    cable type polyurethane   [1]
    notes
    [1] two-wire shielded cable with twisted pair Wire    
     
    [2] typically    
     
    [3] zero-based balancing line, determined using the method of least squares    
     
    [4] no equivalent in SI units    
     
    [5] the upper cut-off frequency tolerance: ±10%    
     
    [6] 1 g = 9.81 m/s²    
     
    [7] details: see PCBs Declaration of conformance PS023 and PS060    
    Accessories
    081A40 mounting bolts with 1/4 "-28 thread -1  
     
    ICS-2 single point calibration at 100 Hz (traceable to NIST) -1  
    (M) metric mounting
    equipment      
     
    M081A61 metric mounting bolts with M6 thread -1  
    (TO) temperature output
    temperature range 36 … 250 ° F 2 … 121 ° C  
     
    Transcription factor 5.56 mV / ° F
    (zero point at 32 °F)
    10 mV / ° C  
     
    connections (red wire) acceleration output    
     
    (Black wire) pin assigNment mass    
     
    pin assigNment (white wire) temperature output    
     
    Pin assigNment (green wire) ground    
    (EX) with ATEX registration
    approval for use in potentially explosive atmospheres ExIA IIC T4, AExia IIC, T4    
     
    approval for use in potentially explosive atmospheres EEx nL IIC T4,-40 ° C≤Ta≤121 ° C, II 3 G    
     
    Approval for use in potentially explosive atmospheres EEx nL IIC T4,-40 ° C≤Ta≤121 ° C, II 1 G    
     
    approval for potentially explosive atmospheres CL I, div I, groups A, B, C, D CL II, div I, groups E, F, G Cl III, div I    
     
    approval for use in potentially explosive atmospheres CL I, div 2 , Groups A, B, C, D ExnL IIC T4, AExnA IIC T4    
    PCB-(M)603C11

    Specific References

    Download

    603C11_drawing.PDF

    Technical drawing

    Download (51.79k)

    603C11_manual.PDF

    Operating instructions

    Download (887.73k)

    603C11_specs.PDF

    Technical data sheet

    Download (89.63k)