  • Industrial ICP® vibration sensor,
  • Measuring range 10 g,
  • Sensitivity500 mV / g,
  • Frequency range 0.17 … 10,000 Hz,
  • One-point calibration at 100 Hz,
  • M12 connection socket top
  •  

    basic data
    sensitivity (±20%) 500 mV / g 51 mV/(m/s²) [1]
     
    measuring range ± 10 g ±98 m / s²  
     
    Frequency range (±5%) 28 … 240,000 cpm 0.47 … 4,000 Hz [4]
     
    frequency range (± 10%) 20 … 300,000 cpm 0.33 … 5,000 Hz  
     
    frequency range (± 3 dB) 10 … 600,000 cpm 0.17 … 10,000 Hz  
     
    resonance frequency 960 kcpm 16,000 Hz [2]
     
    wide band resolution (1 … 10,000 Hz) 35 µg 343 µm / s ² [2]
     
    linearity ≤1%   [3]
     
    Transverse acceleration sensitivity ≤ 7%    
    conditions of use
    (shock) overload limit 49,050 m/s ² pk pk 5,000 g  
     
    Temperature range -65 … 250 ° F -54 … 121 ° C  
     
    temperature impact on Temperature coefficient of sensitivity see curve below   [2]
     
    Protection class IP68    
    technical characteristics
    settling time (on area within ±1% of the bias voltage) ? 10s    
     
    Discharging charging time constant ≥ 1.0 s    
     
    power 18 … 28 V DC    
     
    Constant power supply 2 … 20 mA    
     
    output impedance < 150 Ω    
     
    bias voltage 8 … 12 V DC    
     
    spectral noise (10 Hz) 3 µg / √Hz 29.4 (µm / s ²) / √Hz [2]
     
    Spectral noise (100 Hz) 0.7 µg / √Hz 6.9 (µm / s ²) / √Hz [2]
     
    spectral noise (1,000 Hz) 0,5 µg / √Hz 4.9 (µm / s ²) / √Hz [2]
     
    electrical insulation    
    physical properties
    size - hex 7/8 in 22.2 mm  
     
    Size - height 1.82 in 46.3 mm  
     
    weight 2.8 oz 80 g  
     
    mounting thread 1/4 "internal thread No metric equivalent [5]
     
    Assembly torque 2 … 5 ft-lb 2.7 … 6.8 Nm  
     
    sensor element ceramic    
     
    geometry of the Sensor element shear principle    
     
    housing material stainless steel (316 L)    
     
    sealing hermetically sealed welded    
     
    Connection top    
    notes
    [1] conversion factor 1 g = 9.81 m/s²    
     
    [2] Typically    
     
    [3] zero-based balancing line, determined using the method of least squares    
     
    [4] the upper cut-off frequency tolerance: ±10%    
     
    [5] no equivalent in SI units    
     
    [6] Details: see PCBs Declaration of conformance PS023    
    Equipment
    081A40 mounting bolts with 1/4 "-28 thread -1  
     
    ICS-2 single point calibration at 100 Hz (Rü" ckführbar on NIST) -1  
    (M) metric mounting
    equipment      
     
    M081A61 metric mounting bolts with M6-thread -1  
    PCB-(M)601A92