PCB-(M)608A11/050BZ
| basic data | |||
| sensitivity (±15%) | 100 mV/g | 10.2 mV/(m/s²) | [3] |
| measuring range | ±50 g | ±490 m / s² | |
| Frequency range (± 3 dB) | 30 … 600,000 cpm | 0.5 … 10,000 Hz | |
| resonance frequency | 1,320 kcpm | 22,000 Hz | [4] |
| Wide band resolution | 350 µg | 3.434 µm / s ² | [4] |
| linearity | ≤1% | [5] | |
| Lateral acceleration sensitivity | ≤ 7% | ||
| Conditions of use | |||
| Overload limit (shock) | 5,000 g pk | 49,050 m/s² pk | |
| Temperature range | -65 … 250 ° F | -54 … 121 ° C | |
| protection class | IP68 | ||
| technical characteristics | |||
| Settling time (on bias voltage ± 1%) | ≤ 2.0 s | ||
| discharging charging time constant | ≥ 0.3 s | ||
| Power | 18 … 28 V DC | ||
| constant power supply | 2 … 20 mA | ||
| Output impedance | < 150 Ω | ||
| Bias voltage | 8 … 12 V DC | ||
| Spectral noise (10 Hz) | 8 µg / √ Hz | 78.5 (µm / s ²) / √ Hz | [4] |
| spectral noise (100 Hz) | 5 µg / √ Hz | 49.1 (µm / s ²) / √ Hz | [4] |
| spectral noise (1,000 Hz) | 4 µg / √ Hz | 39.2 (µm / s ²) / √ Hz | [4] |
| electric Insulation (housing) | ≥ 100 MΩ | ||
| physical properties | |||
| size - Allen | 9/16 in | 14 mm | |
| size ße - height | 2.5 in | 64 mm | |
| weight (with cord) | 3.5 oz | 99.3 grams | |
| fixing | mounting bolt | ||
| Mounting thread | 1/4 "-28 threaded | [1] | |
| Assembly torque | 2 … 5 ft-lb | 2.7 … 6.8 Nm | |
| sensor element | " Ceramic | ||
| geometry of the sensor element | shear principle | ||
| housing material | stainless steel | ||
| Sealing | shed | ||
| electrical connection | potted integrated cable | ||
| cable outlet | top | ||
| Cable length | 50 ft | 15.2 m | |
| cable type | polyurethane | [2] | |
| notes | |||
| [1] | No equivalent in SI units | ||
| [2] | Screened two-wire cable with twisted wire | ||
| [3] | conversion factor 1 g = 9.81 m/s² | ||
| [4] | typically | ||
| [5] | zero-based balancing line, determined using the method of least squares | ||
| [6] | details: see PCBs Declaration of conformance PS023 and PS060 | ||
| Equipment | |||
| 081A40 | mounting bolts with 1/4 "-28 thread | -1 | |
| ICS-2 | single point calibration at 100 Hz (Rü" ckführbar on NIST) | -1 | |
| (M) metric mounting | |||
| equipment | |||
| M081A61 | metric mounting bolts with M6-thread | -1 | |
| (TO) temperature output | |||
| temperature range | 36 … 250 ° F | 2 … 121 ° C | |
| transcription factor | 5.56 mV / ° F (zero point at 32 °F) | 10 mV / ° C | |
| Electrical connection | potted integrated cable | ||
| connections (red wire) | acceleration output | ||
| (Black wire) pin assigNment | mass | ||
| pin assigNment (white wire) | temperature output | ||
| (EX) with ATEX approval | |||
| Admission f &kitchen r hazardous atmospheres | CL I, div 2, groups A, B, C, D ExnL IIC T4, AExnA IIC T4 | ||
| approval for use in potentially explosive atmospheres | EEx nL IIC T4, 40 ° C ≤ Ta ≤ 121 ° C, II 3 G | ||
| approval for use in potentially explosive atmospheres | EEx nL IIC T4,-40 ° C ≤ Ta ≤ 121 ° C, II 1 G | ||
| Approved for use in potentially explosive atmospheres | CL I, div I, groups A, B, C, D CL II, div I, groups E, F, G Cl III, div I | ||
| approval for hazardous Atmospheres | ExIA IIC T4, AExia IIC, T4 | ||
| (LB) low bias voltage | |||
| Bias stress | 4.8 … 6.5 V DC | ||
| Power | 12 … 28 V DC | ||
| measuring range | ± 10 g | ±98 m / s ² |