PCB-(M)350D02
- ICP® shock acceleration sensor,
- Masseisoliert,
- With electric and mechanical filter,
- Measuring range 50,000 g,
- Sensitivity 0.1 mV / g,
- Frequency range 4 … 10,000 Hz,
- Integrated cable, length 3 m, ending on 10-32 plug
| basic data | |||
| Sensitivity (± 30%) | 0.1 mV / g | 0.01 mV / (m / s) | |
| measuring range | ± 50,000 g pk | ± 490,000 m / spk | |
| Frequency range (± 1 dB) | 4 ... 10,000 Hz | ||
| Frequency range (-3 dB) | 2 ... 25,000 Hz | [4] | |
| Corner frequency of the electric filter (-3 dB) | 17,000 Hz | [1] [2] | |
| Resonance frequency of the mechanical filter | 45,000 Hz | [1] [3] | |
| resonant frequency | ≥100.000 Hz | ||
| Broadband resolution | 0.5 g RMS | 4.9 m / s RMS | [1] |
| Linearity error (per 10,000 g / 98,100 m / s²) | ≤2,5% | ||
| Lateral acceleration sensitivity | ≥7% | ||
| Operating conditions | |||
| Overload limit (shock) | ± 150,000 g pk | 1,471,500 ± m / s² pk | |
| The temperature range (operation) | 0 ... 150 ° F | -18 ... 66 ° C | |
| Temperature range (storage) | -40 ... 200 ° F | -40 ... 93 ° C | |
| Technical characteristics | |||
| supply voltage | 20 ... 30 V DC | ||
| Constant current supply | 2 ... 20 mA | ||
| output impedance | ≤200 Ω | ||
| Bias voltage | 8 ... 14 V DC | ||
| discharging | 0.10 s | [1] | |
| Settling time (on the bias voltage ± 10%) | <1 s | ||
| Electrical insulation (housing) | > 1 M | ||
| physical properties | |||
| sensor element | ceramics | ||
| Geometry of the sensor element | shear principle | ||
| housing material | titanium | ||
| seal | hermetically sealed | ||
| Size - Hexagon | 0.375 in | 9.5 mm | |
| Size - Height | 0.87 in | 22.1 mm | |
| Weight (without cable) | 0.16 oz | 4.5 grams | [1] |
| electrical connection | integrated cable | ||
| cable length | 10 ft | 3 m | |
| cable type | Type 031, twisted pair | ||
| mounting thread | 1/4 "-28 male thread | ||
| Hints | |||
| [1] | typically, | ||
| [2] | second-order low-pass filter | ||
| [3] | The amplitude is increased by 9 dB in the range of the resonance frequency. | ||
| [4] | When given upper frequency limit is a guideline. The frequency response is limited by the combination of electrical and mechanical low pass filter. | ||
| included supply | |||
| ACS-14 | Hopkinson bar calibration for shock sensor | -1 | |
| ACS-22 | Frequency response, traceable to NIST (100 Hz to ± 1 dB point) | -1 | |
| (M) Metric assembly | |||
| mounting thread | M6 x 0.75 male |