PCB-123A
basic data | |||
sensitivity (±20%) | 1.1 pC/psi | 0,159 pC/kPa | |
measuring range | 3,000 psi | 20.685 kPa | |
Maximum pressure | of5,000 psi | 34.475 kPa | |
resolution | 0,02 psi | 0.138 kPa | [1] |
resonance frequency | ≥ 25,000 Hz | [2] | |
rise time (reflected pressure) | &le≤ 20 µs | [2] | |
linearity | ≤ 1% FS | [3] | |
Conditions of use | |||
Acceleration sensitivity | ≤ 0.002 psi/g | ≤ 1,41E-3 kPa/(m/s²) | |
Temperature range (operation) | -450 … 500 ° F | -268 … 260 ° C | |
Temperature coefficient of sensitivity | ≤0,01 %/°F | ≤ 0.018 %/°C | |
maximum shock (for fraction of a second) | 10,000 ° F | 5.538 ° C | |
maximum shock resistance | 10,000 g PK | 98.070 m /PK | |
Technical features | |||
output polarity (positive pressure) | negative | ||
capacity | 18 pF | ||
Electrical insulation | ≥100 MΩ | ||
insulation resistance (at room temperature) | ≥ 1,000,000 M Ω | ||
-physical Properties | |||
sensor element | quartz | ||
geometry of the sensor element | compression principle | ||
parts usematerial | 17-4 (Stainless steel) | ||
membrane | INVAR | ||
waterproofing | epoxy resin | ||
Electrical connection | integrated cable | ||
Cable length | 4 ft | 1.2 m | |
helium flow velocity (1,500 psi) | 40 SCF/h | 1.13 m ³/h | |
Water flow rate (at 50 psi) | /1.2 gal / min | 4.5 l/min | |
mass | 4.23 oz | 120 grams | |
Notes | |||
[1] | resolution depending on the cable length and the range setting of the signal processing | ||
[2] | with helium flow | ||
[3] | zero-based balancing line, determined using the method of least squares | ||
Accessories | |||
065A15 | seal, 0,624 "outside x 0,553" x 0.080 ", brass | 3 | |
070A08 | cable adapter (10 32 female to BNC female ) | -1 |