PCB-(M)102B15
zoom_out_map

PCB-(M)102B15

  • ICP® pressure sensor with INVAR-membrane,
  • Masseisoliert,
  • Beschleunigungskompensiert,
  • Measuring range 1.379 kPa,
  • Sensitivity 3.6 mV / kPa,
  • Lower limiting frequency 0.5 Hz,
  • 10-32 female connector
  •  

    basic data
    (±5 V output range) measuring range 1,379 kPa  
     
    usable range (output range ±10 V) 400 psi 2.758 kPa [3]
     
    sensitivity (± 5 mV/psi) 25 mV / psi 3.6 mV / kPa  
     
    maximum static pressure 1,000 psi 6.895 kPa  
     
    Resolution 0.001 psi 0.007 kPa [1]
     
    resonance frequency ≥ 500, 000 Hz    
     
    rise time (reflected pressure) ≤ 1.0 µs    
     
    Lower limiting frequency (-5%) 0.5 Hz    
     
    linearity ≤1.0% FS   [2]
    Conditions of use
    acceleration sensitivity ≤ 0.002 psi/g ≤ 0,0014 kPa/(m/s²)  
     
    temperature range (operation) -100 … 275 ° F -73 … 135 ° C  
     
    Temperature coefficient of sensitivity ≤ 0.03 %/°F ≤ 0,054 %/°C [1]
     
    maximum temperature shock(for fraction of a second) 3,000 ° F 1,650 ° C  
     
    Maximum shock resistance 20,000 g pk 196,000 m/s² pk  
    Technical characteristics
    Output polarity (positive pressure) positive    
     
    Discharging charging time constant (at room temperature) ≥ 1.0 s    
     
    supply 20 … 30 V DC    
     
    Constant power supply 2 … 20 mA    
     
    output impedance < 100Ω    
     
    Bias voltage 8 … 14 V DC    
     
    electrical insulation 100 MΩ  
    -physical Properties
    geometry of the sensor element compression principle    
     
    sensor element quartz    
     
    parts usematerial stainless steel    
     
    membrane INVAR    
     
    sealed sealeddensely welded    
     
    electrical connection 10-32-coaxial socket    
     
    mass 0.41 oz 11.6 grams  
    Notes
    [1] typically    
     
    [2] zero-based balancing line, determined using the method of least squares    
     
    [3] for output range up to 10 V, the supply voltage must be at least 24 V DC. At low bias voltage of the sensor, no negative output range is usable up to-10 V.    
     
    [4] details: see PCBs Declaration of conformance PS023    
    equipment
    065A03 sealing ring, 0.435 "au need x 0,377 "x 0.030" high brass 3  
    (M) metric mounting
    equipment      
     
    065A40 Seal ring, 0.435 "outside x 0,397" x 0.030 "high brass 3  
    (S) stainless steel diaphragm
           