PCB-(M)102B15
    zoom_out_map

      PCB-(M)102B15

    • ICP® pressure sensor with INVAR-membrane,
    • Masseisoliert,
    • Beschleunigungskompensiert,
    • Measuring range 1.379 kPa,
    • Sensitivity 3.6 mV / kPa,
    • Lower limiting frequency 0.5 Hz,
    • 10-32 female connector
    •  

      basic data
      (±5 V output range) measuring range1,379 kPa 
       
      usable range (output range ±10 V)400 psi2.758 kPa[3]
       
      sensitivity (± 5 mV/psi)25 mV / psi3.6 mV / kPa 
       
      maximum static pressure1,000 psi6.895 kPa 
       
      Resolution0.001 psi0.007 kPa[1]
       
      resonance frequency≥ 500, 000 Hz  
       
      rise time (reflected pressure)≤ 1.0 µs  
       
      Lower limiting frequency (-5%)0.5 Hz  
       
      linearity≤1.0% FS [2]
      Conditions of use
      acceleration sensitivity≤ 0.002 psi/g≤ 0,0014 kPa/(m/s²) 
       
      temperature range (operation)-100 … 275 ° F-73 … 135 ° C 
       
      Temperature coefficient of sensitivity≤ 0.03 %/°F≤ 0,054 %/°C[1]
       
      maximum temperature shock(for fraction of a second)3,000 ° F1,650 ° C 
       
      Maximum shock resistance20,000 g pk196,000 m/s² pk 
      Technical characteristics
      Output polarity (positive pressure)positive  
       
      Discharging charging time constant (at room temperature)≥ 1.0 s  
       
      supply20 … 30 V DC  
       
      Constant power supply2 … 20 mA  
       
      output impedance< 100Ω  
       
      Bias voltage8 … 14 V DC  
       
      electrical insulation100 MΩ 
      -physical Properties
      geometry of the sensor elementcompression principle  
       
      sensor elementquartz  
       
      parts usematerialstainless steel  
       
      membraneINVAR  
       
      sealedsealeddensely welded  
       
      electrical connection10-32-coaxial socket  
       
      mass0.41 oz11.6 grams 
      Notes
      [1]typically  
       
      [2]zero-based balancing line, determined using the method of least squares  
       
      [3]for output range up to 10 V, the supply voltage must be at least 24 V DC. At low bias voltage of the sensor, no negative output range is usable up to-10 V.  
       
      [4]details: see PCBs Declaration of conformance PS023  
      equipment
      065A03sealing ring, 0.435 "au need x 0,377 "x 0.030" high brass3 
      (M) metric mounting
      equipment   
       
      065A40Seal ring, 0.435 "outside x 0,397" x 0.030 "high brass3 
      (S) stainless steel diaphragm
          
      PCB-(M)102B15

      Specific References

      Download

      102B15_drawing.PDF

      Technical drawing

      Download (78.87k)

      102B15_manual.PDF

      Operating instructions

      Download (424.69k)

      102B15_specs.PDF

      Technical data sheet

      Download (252.38k)