PCB-EX(M)608A11
basic data | |||
sensitivity (±15%) | 100 mV/g | 10.2 mV/(m/s²) | [3] |
measuring range | ±50 g | ±490 m / s² | |
Frequency range (± 3 dB) | 30 … 600,000 cpm | 0.5 … 10,000 Hz | |
resonance frequency | 1,320 kcpm | 22,000 Hz | [4] |
Wide band resolution | 350 µg | 3.434 µm / s ² | [4] |
linearity | ≤1% | [5] | |
Lateral acceleration sensitivity | ≤ 7% | ||
Conditions of use | |||
Overload limit (shock) | 5,000 g pk | 49,050 m/s² pk | |
Temperature range | -65 … 250 ° F | -54 … 121 ° C | |
protection class | IP68 | ||
approval for potentially explosive atmospheres | CL I, div 2, groups A, B, C, D ExnL IIC T4, AExnA IIC T4 | ||
approval for hazardousAtmospheres | EEx nL IIC T4,-40 ° C ≤ Ta ≤ 121 ° C, II 3 G | ||
approval for use in potentially explosive atmospheres | EEx nL IIC T4,-40 ° C ≤ Ta ≤ 121 ° C, II 1 G | ||
Approval for use in potentially explosive atmospheres | CL I, div I, groups A, B, C, D CL II, div I, groups E, F, G Cl III, div I | ||
registration for use in potentially explosive atmospheres | ExIA IIC T4, AExia IIC, T4 | ||
technical characteristics | |||
settling time (on bias voltage ± 1%) | ≤ 2.0 s | ||
Discharging charging time constant | ≥ 0.3 s | ||
power | 18 … 28 V DC | ||
Constant power supply | 2 … 20 mA | ||
output impedance | < 150 Ω | ||
Bias voltage | 8 … 12 V DC | ||
spectral noise (10 Hz) | 8 µg / √ Hz | 78.5 (µm / s ²) / √ Hz | [4] |
Spectral noise (100 Hz) | 5 µg / √ Hz | 49,1 (µm / s ²) / √ Hz | [4] |
spectral noise (1,000 Hz) | 4 µg / √ Hz | 39.2 ( µm / s ²) / √ Hz | [4] |
electrical insulation (housing) | ≥ 100 MΩ | ||
physical properties | |||
size -. Allen | 9/16 in | 14 mm | |
size - height | 2.5 in | 64 mm | |
weight (with cord) | 3.5 oz | 99.3 grams | |
Fixing | mounting bolt | ||
mounting thread | 1/4 "-28 threaded | [1] | |
" Mounting torque | 2 … 5 ft-lb | 2.7 … 6.8 Nm | |
sensor element | ceramic | ||
geometry of the sensor element | Shear principle | ||
housing material | stainless steel | ||
sealing | shed | ||
Electrical connection | spilled integratedCable | ||
cable outlet | top | ||
cable length | 10 ft | 3 m | |
cable type | Polyurethane | [2] | |
notes | |||
[1] | no equivalent in SI units | ||
[2] | two-wire shielded cable with twisted pair wires | ||
[3] | conversion factor 1 g = 9.81 m/s² | ||
[4] | typically | ||
[5] | zero-based balancing line, determined using the method of least squares | ||
[6] | details: see PCBs Declaration of conformance PS023 and PS060 | ||
Equipment | |||
081A40 | mounting bolts with 1/4 "-28 thread | -1 | |
ICS-2 | single point calibration at 100 Hz (Rü" ckführbar on NIST) | -1 | |
(M) metric mounting | |||
equipment | |||
M081A61 | metric mounting boltswith M6 threaded | -1 |