PCB-(M)354C10
    zoom_out_map

      PCB-(M)354C10

    • Triaxial ICP® miniature vibration sensor,
    • With through hole,
    • Masseisoliert,
    • Measuring range 500 g
    • Sensitivity 10 mV / g,
    • Frequency range 2 … 8,000 Hz,
    • Mass 5 grams,
    • Integrated connection cable, length 1.5 m,
      ending on 4-pin female connector,
    • including extension cable, length 1.5 m,
      ending to three BNC
    •  

      basic data
      sensitivity (±15%)10 mV / g1.02 mV/(m/s²) 
       
      range±500 g pk ±4. 905 m/s 2 pk 
       
      frequency range (±5%)2 … 8,000 Hz  
       
      Resonance frequency≥40,000 Hz  
       
      Wide band resolution0.003 g RMS0.03 m/s² RMS[2]
       
      Linearity error≤1% [3]
       
      transverse acceleration sensitivity≤5%  
      conditions
      (shock) overload limit± 10.000 g PK±98.100 m/s² pk 
       
      Temperature range-65 … 250 ° F-54 … 121 ° C[1]
      specifications
      power supply18 … 30 V DC  
       
      Constant power supply2 … 20 mA  
       
      output impedance≤200Ω  
       
      Bias voltage7 … 11 V DC  
       
      Discharging charging time constant0.3 … 1,0 s  
       
      settlingbr(auf Bias-Spannung ±10 %)< 3 s  
       
      Spectral noise (1 Hz)1,000 µg / √Hz9.810 (µm / s ²) / √Hz[2]
       
      spectral noise (10 Hz)300 µg / √Hz2.943 ( µm / s ²) / √ Hz[2]
       
      spectral noise (100 Hz)80 µg / √Hz785 (µm / s ²) / √Hz[2]
       
      spectral noise ( 1,000 Hz)25 µg / √Hz308 (µm / s ²) / √Hz[2]
       
      electrical insulation  
      physical properties
      Sensor elementceramic  
       
      geometry of the sensor elementshear principle  
       
      Housing materialtitanium  
       
      sealinghermetically sealed  
       
      size - H öhe0.30 in7.6 mm 
       
      size - length0.55 in14.0 mm 
       
      size - width0.55 in14.0 mm 
       
      Dimensions (without cable)0.18 oz5.0 grams[2]
       
      wiring1/4 "-28 4-pin  
       
      Cable outletOn the side  
       
      Cable length5 ft1.5 m 
       
      cord typetype 034, 4-wire shielded cable  
       
      Fixinghole  
      notes
      [1]the data for temperatures above 121 ° C applies only for transducers with HT option.  
       
      [2]typically  
       
      [3]zero-based balancing line, determined using the method of least squares  
       
      [4]Details: see PCBs Declaration of conformance PS023  
      Accessories
      034 G 054-wire shielded cable, 1.5 m, 4-pin connector to three BNC-1 
       
      039A21Allen wrench 3/32 "hex-1 
       
      081 B 93insulated mounting screw, 5-40 thread x 12.7 mm long (for model 354 C 10)-1 
       
      ACS-1Tfrequency response for Triaxial accelerometer, traceable to NIST, 10 Hz up to the upper 5% frequency-1 
      (M) Metric mounting
      equipment   
       
      M039A22Allen wrench for use with thread M3-1 
       
      M081B93metric mounting screw-1 
      (HT) high temperature electronics
      temperature range-65 … 325 ° F-54 … 163 ° C 
      PCB-(M)354C10