PCB-352C43
    zoom_out_map

      PCB-352C43

    • ICP®-miniature vibration sensor,
    • Masseisoliert,
    • Measuring range 500 g
    • Sensitivity 10 mV / g,
    • Frequency range 1 … 8,000 Hz,
    • Mass of 3 grams,
    • 10-32 female connector above
    • basic data
      sensitivity (±10%)10 mV / g1.02 mV/(m/s²)
      range±500 g pk ±4. 905 m/s 2 pk
      frequency range (±5%)1 … 8,000 Hz
      Frequency range (± 10%)0.5 … 10,000 Hz
      frequency range (± 3 dB)0.3 … 13,000 Hz
      Resonance frequency≥ 30,000 Hz
      wide band resolution0.0008 g RMS0.008 m/s² RMS[3]
      Linearity≤ 1%[1]
      transverse acceleration sensitivity≤ 5%[2]
      conditions of use
      Overload limit±5. 000 g pk±49. 050 m/s² pk
      Temperature range-65… 250 ° F-54 … 121 ° C
      specifications
      power supply22 … 30 V DC
      Constant power supply2 … 20 mA
      output impedance≤200Ω
      bias voltage8 … 15 V DC
      discharging charging time constant0.5 … 2.0 s
      Settling time < br(auf Bias-Spannung ±10 %)< 10 s
      spectral noise (1 Hz)200 µg / √ Hz1,962 (µm / s ²) / √ Hz[3]
      Spectral noise (10 Hz)60 µg / √ Hz589 (µm / s ²) / √ Hz[3]
      spectral noise (100 Hz)15 µg / √ Hz147 (µm / s ²) / √ Hz[3]
      spectral noise (1,000 Hz)8 µg / √ Hz79 (µm / s ²) / √ Hz[3]
      electrical insulation≥ 100 MΩΩ
      Physical properties
      size - height0.42 in10.7 mm
      dimensions0.10 oz3.0 grams[3]
      Sensor elementceramic
      Size - hex0,438 in11.1 mm
      geometry of the sensor elementshear principle
      parts usematerialtitanium
      sealinghermetically sealed welded
      electrical connection10-32-coaxial socket
      Connection positionabove
      attachmentbonding
      notes
      [1]zero-based balancing line, determined using the method of least squares
      [2]transverse acceleration sensitivity typically ≤ 3%
      [3]typically
      [4]details: see PCBs Declaration of conformance PS023
      Accessories
      080A109mounting wax-1
      080A90glue-1
      ACS-1Frequency response, traceable to NIST (10 Hz up to the upper 5% point)-1
      PCB-352C43

      Specific References

      Download

      352C43_drawing.PDF

      Technical drawing

      Download (61.51k)

      352C43_manual.PDF

      Operating instructions

      Download (1.68M)

      352C43_specs.PDF

      Technical data sheet

      Download (73.48k)