PCB-(M)102A05
    zoom_out_map

      PCB-(M)102A05

    • ICP® pressure sensor with INVAR-membrane,
    • Masseisoliert,
    • Beschleunigungskompensiert,
    • Measuring range 690 kPa,
    • Sensitivity 7.3 mV / kPa,
    • Lower limiting frequency 0.5 Hz,
    • 10-32 female connector
    •  

      basic data
      measuring range (±5 V output range)100 psi690 kPa 
       
      usable range (output range ±10 V)1,379 kPa[2]
       
      Sensitivity (± 10 mV/psi)50 mV / psi7.3 mV / kPa 
       
      maximum static pressure1,000 psi6,900 kPa 
       
      Resolution0.002 psi0.014 kPa 
       
      Resonance frequency≥250,000 Hz  
       
      rise time (reflected pressure)≤ 2.0 µs  
       
      Lower cut-off frequency (-5%)0.5 Hz  
       
      Linearity≤1.0% FS [1]
      conditions
      acceleration sensitivity≤ 0.002 psi/g≤ 0,0014 kPa/(m/s²) 
       
      Temperature range (operation)-100 … 275 ° F-73 … 135 ° C 
       
      Temperature coefficient of sensitivity≤ 0.03 %/°F≤ 0,054 %/°C 
       
      maximum shock (f & uuml; r fraction of a second)3,000 ° F1,650 ° C 
       
      maximum shock resistance20,000 g pk196,000 m/s² PK 
      Technical features
      output polarity (positive pressure)positive  
       
      discharging charging time constant (at room temperature)≥ 1.0 s  
       
      Supply20 … 30 V DC  
       
      constant power supply2 … 20 mA  
       
      output impedance< 100Ω  
       
      Bias voltage8 … 14 V DC  
       
      Electrical insulation100 MΩ 
      physical properties
      geometry of the sensor elementcompression principle  
       
      Sensor elementquartz  
       
      housing materialstainless steel  
       
      membraneINVAR  
       
      Sealinghermetically sealed welded  
       
      electrical connection10-32-coaxial socket  
       
      Mass0.41 oz11.6 grams 
      notes
      [1]zero-based balancing line, determined using the method of least squares  
       
      [2]for output range supply voltage up to 10 V must be at least 24 V DC. At low bias voltage of the sensor, no negative output range is usable up to-10 V.  
       
      [3]details: see PCBs Declaration of conformance PS023  
      equipment
      065A03sealing ring, 0.435 "au need x 0,377 "x 0.030" high brass3 
      (M) metric mounting
      equipment   
       
      065A40Seal ring, 0.435 "outside x 0,397" x 0.030 "high brass3 
      (S) stainless steel diaphragm
      diaphragm316 L (stainless steel)  
      PCB-(M)102A05

      Specific References

      Download

      102A05_drawing.PDF

      Technical drawing

      Download (124.9k)

      102A05_manual.PDF

      Operating instructions

      Download (676.16k)

      102A05_specs.PDF

      Technical data sheet

      Download (427.83k)