PCB-352B70
,
basic data | |||
sensitivity (±25%) | 1 mV / g | 0.1 mV/(m/s²) | |
range | ±5. 000 g pk | ±49. 000 m/s² pk | |
frequency range (±5%) | 0.7 … 9,000 Hz | ||
Frequency range (-3 dB) | 0.4 … 20,000 Hz | ||
integrated filter | low-pass | ||
electric filter corner frequency | 23,000 Hz | [1] | |
the filter slope | 12 dB/octave | ||
resonance frequency | ≥55,000 Hz | ||
Wide band resolution | 0.025 g RMS | 0.25 m/s² RMS | [1] |
Linearity error | ≤1% | [2] | |
transverse acceleration sensitivity | ≤5% | ||
conditions | |||
(shock) overload limit | ± 10.000g pk | ±98. 000 m/s² pk | |
temperature range (operation) | -65 … 250 ° F | -54 … 121 ° C | |
specifications | |||
power supply | 20 … 30 V DC | ||
Constant power supply | 2 … 20 mA | ||
output impedance | ≤200Ω | ||
Bias voltage | 8 … 14 V DC | ||
discharging charging time constant | 0.5 … 1 | ||
settling time (on Bias-Spannung ±10%) | <5 s | ||
spectral noise (1 Hz) | 7,200 µg / √Hz | 70.600 (µm / s ²) / √Hz | [1] |
Spectral noise (10 Hz) | 2,400 µg / √Hz | 23,500 (µm / s ²) / √Hz | [1] |
spectral noise (100 Hz) | 740 µg / √Hz | 7.250 (µm / s ²) /. ? Hz | [1] |
Spectral noise (1,000 Hz) | 190 µg / √Hz | 1,860 (µm / s ²) / √Hz | [1] |
electrical insulation (base) | > 100,000,000Ω | ||
Physical properties | |||
size - height | 0.90 in | 22.9 mm | |
size - hex | 3/8 in | 9.5 mm | |
Mass | 0.15 oz | 4.3 grams | [1] |
sensor element | ceramic | ||
geometry of the sensor element | Shear principle | ||
housing material | titanium | ||
sealing | hermetic | ||
Electrical connection | 10-32-coaxial socket | ||
Connection position | above | ||
mounting thread | 10-32 female thread | ||
notes | |||
[1] | Typically | ||
[2] | zero-based balancing line, determined using the method of least squares | ||
[3] | Details: see PCBs Declaration of conformance PS023 | ||
equipment | |||
080A04 | adhesive mounting, 0.37 "hex x. 0.2 "thick, 10-32 thread, anodized aluminum | -1 | |
080A109 | mounting wax | -1 | |
081 B 05 | Mounting bolts (10-32 on 10-32) | -1 | |
ACS-23 | frequency response, traceable to NIST (5 Hz … 5% point, equal frequency intervals ±10%) | -1 | |
M081B05 | Mounting bolts 10-32 to M6 x 0.75 | -1 |