PCB-(M)350B21
basic data | |||
sensitivity (±30%) | /0.05 mV / g | 0.005 mV/(m/s²) | |
range | ±100. 000 g PK | ±980. 000 m/s² pk | |
Frequency range (± 1 dB) | 1 … 10,000 Hz | ||
frequency range (± 3 dB) | 0,5 … 35,000 Hz | ||
Resonance frequency | ≥ 200.000 Hz | ||
wide band resolution | 0.3 g RMS | 2,9 m/s² RMS | [1] |
Linearity error (per 10,000 g / 98.100 m/s²) | ±0. 5% | [2] | |
transverse acceleration sensitivity | ≤ 7% | ||
Conditions of use | |||
Overload limit (shock) | ±200. 000 g pk | ±1. 960.000 m/s² pk | |
Temperature range (operation) | -65 …200 °F | -54 … 93 ° C | |
Strain sensitivity (footprint) | 0.2 g/µε | 2.0 (m/s²)/µε | [1] |
technical characteristics | |||
Supply voltage | 18 … 30 V DC | ||
constant power supply | 2 … 20 mA | ||
Output impedance | ≤100 Ω | ||
Bias voltage | 8 … 12 V DC | ||
discharging charging time constant | 0.5 … 0,7 s | [1] | |
Settling time < br(auf Bias-Spannung ±10 %) | <5 s | ||
electrical insulation (housing) | >100 MΩ | ||
Physical properties | |||
Sensor element | ceramic | ||
Geometry of the sensor element | shear principle | ||
housing material | titanium | ||
sealing | Hermetically sealed | ||
size-Hexagonal | 0.375 in | 9.5 mm | |
size - height | 0.73 in | 18.5 mm | |
dimensions (without cable) | 0.15 oz | 4.4 grams | [1] |
Electrical connection | integrated cable | ||
cable length | 10 ft | 3 m | |
cable type | type 031,. twisted pair | ||
mounting thread | 1/4 "-28 thread | ||
notes | |||
[1] | typically | " | |
[2] | maximum sensitivity drift | ||
Accessories | |||
ACS-14 | Hopkinson bar calibration for shock resistance thermometer | -1 | |
ACS-22 | frequency response, traceable to NIST (100 Hz to ±1dB point) | -1 | |
(M) metric mounting | |||
mounting thread | M6 x 0.75 m ännlich |