PCB-352A24

PCB-352A24

  • ICP®-miniature vibration sensor,
  • Mass 0.8 grams,
  • Masseisoliert,
  • Measurement range 50 g,
  • Sensitivity 100 mV/g,
  • Frequency range 1 … 8,000 Hz,
  • 3 56-connector side,
  • Including connection cable, length 3 m, ending on 10-32 plug

basic data
sensitivity (±10%)100 mV/g10.2 mV/(m/s²)
range±50 g pk ±490 m/s 2 pk
frequency range (±5%)1.0 … 8,000 Hz
Frequency range (± 10%)0.8 … 10,000 Hz
frequency range (± 3 dB)0.4 … 12,000 Hz
Resonance frequency≥ 30,000 Hz
wide band resolution0.0002 g RMS0.002 m/s² RMS[1]
Linearity≤ 1%[2]
transverse acceleration sensitivity≤ 5%
conditions of use
Overload limit (shock)±5. 000 g pk±49. 050 m/s² pk
Temperature range (operation)-65 … 250 ° F-54 … 121 ° C
specifications
power supply18 … 30 V DC
constant power supply2 … 20 mA
output impedance≤ 300 Ω
Bias voltage8 … 12 V DC
discharging charging time constant0.4 … 1.5 s
Settling time < br(auf Bias-Spannung ±10 %)< 8 s
spectral noise (1 Hz)80 µg / √ Hz785 (µm / s ²) / √ Hz
Spectral noise (10 Hz)15 µg / √ Hz147 (µm / s ²) / √ Hz[1]
spectral noise (100 Hz)4 µg / √ Hz39 (µm / s ²) / √ Hz[1]
spectral noise (1,000 Hz)1 µg / √ Hz9.8 (µm / s ²) / √ Hz[1]
electric Insulation (base)≥100 MΩ
physical properties
size - height0.19 in4,8 mm
size - length0.48 in12.2 mm
size - width0.28 in7.1 mm
mass0.03 oz0.8 grams[1]
Sensor elementCeramic
Geometry of the sensor elementshear principle
housing materialanodized aluminum
Waterproofingepoxy resin
electrical connection3-56 coaxial plug
connection positionside
Attachmentbonding
notes
[1]typically
[2]zero-based balancing line, determined using the method of least squares
[3]details: see PCBs Declaration of conformance PS023
includedAccessories
030A10coaxial cable, 3 m, 3-56-plug to 10-32-1
039A28transfer tool-1
080A109mounting wax-1
ACS-1frequency response, r ückführbar on NIST (10 Hz up to the upper 5% point)-1
(RH) RoHS compliant
Accessories
RH030A10Cable-1

19 other articles of the same category: