PCB-260M29
- Triaxial ICP® force sensor,
- Measuring range (Z-axis) 0.445 kN,
- Measuring range (X and Y axis) 0.444 kN,
- Sensitivity (Z-axis) 5.60 mV / N,
- Sensitivity (X and Y axis) 11.25 mV / N,
- Frequency range (Z-axis) 0.01 ... 90.000 Hz,
- Frequency range (X and Y axis) 0.001 ... 90,000 Hz,
- 4-pin connection socket on the side
| specifications | |||
|---|---|---|---|
| basic data | English units | SI units | |
| Sensitivity (Z-axis) (± 15%) | 25 mV/lb | 5,60 mV/N | [2][3] |
| Sensitivity (X and Y axis) (± 15%) | 50 mV/lb | 11,25 mV/N | [2][3] |
| Measuring range (Z-axis) | 100 lb | 0,445 kN | |
| Measuring range (X and Y axis) | 100 lb | 0,444 kN | |
| Maximum force (Z-axis) | 1200 lb | 5,33 kN | |
| Maximum force (X and Y axis) | 100 lb | 0,445 kN | |
| Maximum moment (Z-axis) | 14 ft-lb | 18,98 Nm | |
| Maximum moment (X and Y axis) | 13 ft-lb | 17,63 Nm | |
| Broadband resolution (Z-axis) | 0,006 lbRMS | 0,027 NRMS | [1] |
| Broadband resolution (X and Y axis) | 0,002 lbRMS | 0,0089 NRMS | [1] |
| Upper limit frequency | 90.000 Hz | ||
| Lower limit frequency (Z-axis) (-5%) | 0,01 Hz | ||
| Lower cutoff frequency (X and Y axis) (-5%) | 0,001 Hz | ||
| linearity error | ≤1 % FS | [4] | |
| Crosstalk (between X and Y axis) | ±3 % | ||
| Crosstalk (between X / Y axis and Z axis) | ±5 % | ||
| Operating conditions | |||
| temperature range | -65 … 250 °F | -54 … 121 °C | |
| Technical characteristics | |||
| Maximum output signal (Z axis) | ±2,5 VDC | ||
| Maximum output signal (X and Y axis) | ±5 VDC | ||
| Discharge time constant (Z-axis) | ≥50 s | ||
| Discharge time constant (X and Y axis) | ≥500 s | ||
| Supply voltage (all axes) | 20 … 30 VDC | ||
| Constant current supply (all axes) | 2… 20 mA | ||
| output impedance | ≤100 Ω | ||
| Bias voltage | 8 … 14 VDC | ||
| Spectral noise (X and Y axis) (1 Hz) | 0,0003 lb/√Hz | 0,00134 N/√Hz | |
| Spectral noise (X and Y axis) (10 Hz) | 0,00015 lb/√Hz | 0,000668 N/√Hz | |
| Spectral noise (X and Y axis) (100 Hz) | 0,00005 lb/√Hz | 0,000223 N/√Hz | |
| Spectral noise (X and Y axis) (1,000 Hz) | 0,000015 lb/√Hz | 0,0000668 N/√Hz | |
| Spectral noise (Z-axis) (1 Hz) | 0,0015 lb/√Hz | 0,00668 N/√Hz | |
| Spectral noise (Z-axis) (10 Hz) | 0,00055 lb/√Hz | 0,00245 N/√Hz | |
| Spectral noise (Z-axis) (100 Hz) | 0,0002 lb/√Hz | 0,000891 N/√Hz | |
| Spectral noise (Z-axis) (1.000 Hz) | 0,000055 lb/√Hz | 0,000245 N/√Hz | |
| output polarity | positive | ||
| Physical properties | |||
| preload | 56 in-lb | 6,3 Nm | [5][6] |
| Stiffness (Z-axis) | 10 lb/µin | 1,75 kN/µm | [1] |
| Stiffness (X and Y axis) | 4 lb/µin | 0,7 kN/µm | [1] |
| Size - height | 0,390 in | 9,90 mm | |
| Size - length | 1,075 in | 27,3 mm | |
| Size - width | 0,95 in | 24,1 mm | |
| Size - inside diameter | 0,319 in | 8,10 mm | |
| Size - diameter of the measuring surface | 0,680 in | 17,27 mm | |
| Mass | 0,927 oz | 26.27 grams | |
| housing material | stainless steel | ||
| seal | Hermetically sealed | ||
| Electrical connection | 4-Pin | ||
| port location | Laterally | ||
[6]
| Hints | |||
|---|---|---|---|
| [1] | typically, | ||
| [2] | When using the recommended preload. The sensitivities may change by ± 5% depending on the preload used. | ||
| [3] | The specified sensitivities apply only when using the supplied bolts. | ||
| [4] | Zero-based balancing line, determined by the least squares method | ||
| [5] | The preload should be at least ten times the operating range used on the X and Y axis. | ||
| [6] | Details: siehe PCB Declaration of Conformance PS023 | ||
| included supply | ||
|---|---|---|
| 081A70 | mounting bolts | (1) |
| 082B02 | washer | (1) |
| 083A10 | mounting sleeve | (1) |
| Available options | ||||
|---|---|---|---|---|
| (W) Waterproof cable connection | ||||
| Kabelabgang | Laterally | |||
| (M) Metric montage | ||||
| included supply | ||||
| M081A70 | mounting bolts | (1) | ||