PCB-356B41/NC
- Triaxial vibration sensor seat cushion,
- For measurements according to DIN EN 30.326,
- Measuring range 10 g,
- Sensitivity 100 mV/g,
- Frequency range 0.5... 1,000 Hz,
- Integrated cable, length 1.5 m,
ending on 4-pin female connector. - Without extension cable li >
basic data | |||
sensitivity (±10%) | 100 mV/g | 10.2 mV/(m/s²) | |
measuring range | ±10 g pk | ±98 m/s 2 pk | |
Frequency range (±5%) | 0.5... 1,000 Hz | [1] | |
resonance frequency | ≥27. 000 Hz | ||
wide band resolution | 0.0002 g RMS | 0,002 m/s² RMS | [3] |
linearity | ≤1% | [2] | |
transverse acceleration sensitivity | ≤5% | ||
conditions of use | |||
Overload limit (shock) | ±2. 000 g pk | ±19. 620 m/s² pk | |
temperature | 14... 122 ° F | 10... 50 ° C | |
temperature effect | ≤ 0.10 %/°F | < 0.17 %/°C | [3] |
specifications | |||
power supply | 6.5... 30 V DC | ||
constant power supply | 0.3... 10 mA | ||
output impedance | ≤500 Ω | ||
Bias voltage | 2.8... 4,5 V DC | ||
discharging charging time constant | 1... 3 s | ||
settling time (auf Bias-Spannung ±10 %) | >20 s | ||
spectral noise (1 Hz) | 50 µg/√Hz | 490 (µm/s²)/√Hz | [] 3] |
spectral noise (10 Hz) | 10 µg/√Hz | 98 (µm/s²)/√Hz | [3] |
spectral noise (100 Hz) | 4 µg/√Hz | 39 (µm/s²)/√Hz | [3] |
Spectral noise (1,000 Hz) | 2 µg/√Hz | 20 (µm/s²)/√Hz | [3] |
physical properties | |||
sensor element | ceramic | ||
Geometry of the sensor element | Shear principle | ||
Housing material | Titanium | ||
Sealing | Hermetic | ||
Size - diameter | 7.87 in | ||
Size - width | 0.472 in | 12 mm | |
Mass | 9.6 oz | 272 grams | [3] |
Electrical connection | Integrated cable | ||
Cable outlet | On the side | ||
Cable termination | 1/4 "-28 4-pin connector | ||
Cable length | 5 ft | 1.5 m | |
Cable type | Type 010, 4-conductor shielded cable | ||
Mounting thread | 5-40 female | ||
Notes | |||
[1] | Standard calibration to 100 Hz with a sensor in the seat cushion. Special calibration about 100 Hz with sensor built in optional. | ||
[2] | Zero-balance just, determined using the method of least squares | ||
[3] | Typically | ||
[4] | Details: see PCBs Declaration of conformance PS023 | ||
Included accessories | |||
081A90 | Mounting bolts, 10-32 on 5-40 | -1 | |
ACS-17 | Calibration of amplitude and phase response, traceable to NIST by 0.5... 100 Hz | -1 |