PCB-350D12
zoom_out_map

PCB-350D12

  • ICP® shock acceleration sensor,
  • Masseisoliert,
  • With electric and mechanical filter,
  • Measuring range 50,000 g,
  • Sensitivity 0.1 mV / g,
  • Frequency range 4 … 10,000 Hz,
  • Integrated cable, length 3 m,
    ending on 10-32 plug
  • "

    basic data
    sensitivity (±30%) 0.1 mV / g 0.01 mV/(m/s²)  
     
    range ±50. 000 g PK ±490. 000 m/s² pk  
     
    Frequency range (± 1 dB) 4 … 10,000 Hz    
     
    frequency range (-3 dB) 2 … 25,000 Hz   [4]
     
    Cutoff frequency of the electrical filter (-3 dB) 17,000 Hz   [1] [2]
     
    resonance frequency of the mechanical filter 45,000 Hz   [1] [3]
     
    Resonance frequency ≥ 100,000 Hz    
     
    Wide band resolution 0.5 g RMS 4.9 m/s 2 RMS [1]
     
    Linearity error (per 10,000 g / 98.100 m/s²) ≤2.5%    
     
    transverse acceleration sensitivity ≤ 7%    
    Conditions of use
    Overload limit (shock) 150.000 g PK ± 1.471.500 m/s² pk  
     
    Temperature range (operation) 0 … 150 ° F -18 … 66 ° C  
     
    temperature range (storage) -40 …200 °F -40 … 93 ° C  
    specifications
    power supply 20 … 30 V DC    
     
    Constant power supply 2 … 20 mA    
     
    output impedance ≤200Ω    
     
    Bias voltage 8 … 14 V DC    
     
    Discharging charging time constant 0.10 s   [1]
     
    settling time < br(auf Bias-Spannung ±10 %) < 1s    
     
    Electrical insulation (housing) > 1 M ?    
    physical properties
    sensor element ceramic    
     
    Geometry of the sensor element Shear principle    
     
    Housing material titanium    
     
    sealed sealedclose    
     
    size - hex 0.375 in 9.5 mm  
     
    size - height 1.04 in 26.5 mm  
     
    Dimensions (without cable) 0.20 oz 5.5 grams [1]
     
    electrical connection integrated cable    
     
    cable length 10 ft 3 m  
     
    Cord type type 032, screened two-core cable with twisted wire    
     
    mounting thread 1/4 "-28 thread    
    Instructions
    [1] typically    
     
    [2] second-order low-pass filter    
     
    [3] the Amplitude is increased in the resonance frequency by 9 dB.    
     
    [4] the specified upper limit frequency is a guideline. The frequency response is limited by the combination of electric and mechanical low pass filter.    
     
    [5] details: see PCBs Declaration of conformance PS023. CE conformity is only ensured if the cable shield(green wire) is grounded.    
    Accessories
    ACS-22 frequency response, traceable to NIST (100 Hz to ±1dB point) -1  
    (M) metric mounting
    Mounting thread M6 x 0.75 male    