PCB-350B44
- Triaxial ICP® shock acceleration sensor
- Ground Isolated
- With electrical and mechanical filter
(-3 dB at 21,000 Hz) - Measuring range 5,000 g
- Sensitivity 1.0 mV / g
- Frequency range 0.4 ... 10,000 Hz
- Integrated connection cable, length 1.5 meters,
ending on 4-pin connection socket - Including extension cable, length 1.5 meters,
ending on three BNC plugs
specifications | |||
---|---|---|---|
basic data | English units | SI units | |
Sensitivity (± 30%) | 1,0 mV/g | 0,1 mV/(m/s²) | |
measuring range | ± 5,000 g pk | ±49.000 m/s²pk | |
Frequency range (± 1 dB) | 0,4 … 10.000 Hz | ||
Frequency range (-3 dB) | 0,2 … 25.000 Hz | [2] | |
resonant frequency | ≥50.000 Hz | ||
Corner frequency of the electrical filter (-3 dB) |
21.000 Hz | [3][1] | |
Mechanical filter resonance frequency | 35.000 Hz | [4][1] | |
Broadband resolution (1 ... 10,000 Hz) | 0,04 gRMS | 0,39 m/s²RMS | [1] |
Linearitätsfehler | ≤2 % | ||
Lateral acceleration sensitivity | ≤7 % | ||
Operating conditions | |||
Overload limit (shock) | ± 50,000 g pk | ±490.000 m/s²pk | |
Temperature range (operation) | -10 … 150 °F | -23 … 66 °C | |
Temperature range (storage) | -40 … 200 °F | -40 … 93 °C | |
Strain sensitivity (base area) | 0.002 g / me | 0.02 (m / s²) / me | [1] |
Technical characteristics | |||
supply voltage | 20 … 30 VDC | ||
Constant current supply | 2 ... 20 mA | ||
output impedance | Ω200 Ω | ||
Bias voltage | 8 … 14 VDC | ||
discharging | 1,0 … 2,0 s | [1] | |
Settling time (on bias voltage ± 10%) |
<10 s | ||
Electrical insulation (base) | >100 MΩ | ||
Physical properties | |||
sensor element | ceramics | ||
Geometry of the sensor element | Scherprinzip | ||
housing material | Titan | ||
seal | Hermetically sealed | ||
Size - height | 1,02 in | 26,0 mm | |
Size - length | 1,02 in | 26,0 mm | |
Size - width | 1,02 in | 26,0 mm | |
Mass (without cable) | 2,2 oz | 62.5 grams | [1] |
Electrical connection | Integrated cable | ||
Kabelabgang | Laterally | ||
cable termination | 4-pin socket (1/4 "-28) | ||
cable length | 5,0 ft | 1,52 m | |
cable type | Type 034, shielded 4-wire cable | ||
attachment | Through holes |
[5]
Hints | |||
---|---|---|---|
[1] | typically, | ||
[2] | The specified upper limit frequency is a guide value. The frequency response is limited by the combination of electrical and mechanical low-pass filters. | ||
[3] | Second order low pass filter | ||
[4] | The amplitude is increased by 9 dB in the range of the resonance frequency. | ||
[5] | Details: siehe PCB Declaration of Conformance PS023 |
included supply | ||
---|---|---|
034G05 | Shielded 4-wire cable, length 1.5 meters, 4-pin connector to three BNC connectors | (1) |
100-16040-40 | Mounting screw M6 x 1.0 | (2) |
100-6774-10 | Mounting screw 1/4 "-28 | (2) |
100-8994-40 | grease | (1) |
ACS-14T | Hopkinson bar calibration for shock transducers | (1) |
ACS-22T | Frequency response for triaxial transducers, traceable to NIST (100 Hz up to the upper ± 1dB point | (1) |