PCB-(M)134A27
- Tourmaline-ICP® pressure sensor,
- For shock waves,
- Measuring range 138,000 kPa,
- Sensitivity 0,04 mV / kPa < br(in Kombination mit Inline-Verstärker 402A17),
- 10-32 socket
| basic data | |||
| Measurement range (output range ± 5 V) | 20,000 psi | 137,900 kPa | |
| Sensitivity (± 15%) | 0.25 mV / psi | 0.04 mV / kPa | [3] |
| Maximum static pressure | 20,000 psi | 137,900 kPa | |
| resolution | 0.3 psi | 2.1 kPa | [2] |
| resonant frequency | ≥1.500.000 Hz | ||
| Rise time (reflected pressure) | ≤0,2 microseconds | ||
| Lower limit frequency (-5%) | 1 Hz | ||
| linearity error | ≤2% FS | [1] | |
| Operating conditions | |||
| The temperature range (operation) | 32 ... 120 ° F | 0 ... 49 ° C | |
| Maximum temperature shock (for split second) | 5000 ° F | 2760 ° C | |
| Technical characteristics | |||
| Output polarity (positive pressure) | positive | ||
| Discharging time (at room temperature) | ≥0,5 s | ||
| supply voltage | 20 ... 30 V DC | ||
| Constant current supply | 2 ... 20 mA | ||
| output impedance | ≤100 Ω | ||
| Bias voltage | 8 ... 14 V DC | ||
| physical properties | |||
| sensor element | tourmaline | ||
| housing material | stainless steel | ||
| membrane | epoxy resin | ||
| seal | epoxy resin | ||
| electrical connection | 10-32 coaxial | ||
| Dimensions | 1.4 oz | 39 grams | |
| Hints | |||
| [1] | Zero-based line of best fit, determined by the method of least squares | ||
| [2] | typically, | ||
| [3] | Calibrated as a system with a 7.5-cm-long cable (Type 003) and an in-line charge amplifier 402A Series | ||
| [4] | Details: see PCB Declaration of Conformance PS023 | ||
| included supply | |||
| 061A30 | 2-pin mounting key | -1 | |
| (M) Metric assembly | |||