  • ICP® pressure sensor with INVAR-membrane,
  • Masseisoliert,
  • Beschleunigungskompensiert,
  • Measuring range 344,7 kPa,
  • Sensitivity 14.5 mV / kPa,
  • Lower limiting frequency 0.5 Hz,
  • 10-32 female connector
  •  

    basic data
    measuring range (±5 V output range) 50 psi 344.7 kPa  
     
    usable range (output range ±10 V) 100 psi 689,4 kPa [1]
     
    sensitivity (±15%) 14.5 mV / kPa  
     
    maximum static pressure 1,000 psi 6.895 kPa  
     
    Resolution 1.0 mpsi 0.007 kPa [2]
     
    resonance frequency ≥ 500, 000 Hz    
     
    rise time (reflected pressure) ≤ 1.0 µs    
     
    lower limiting frequency (-5%) 0.5 Hz    
     
    linearity ≤1.0% FS   [3]
    conditions
    acceleration sensitivity ≤ 0.002 psi/g ≤ 0,0014 kPa/(m/s²)  
     
    temperature range (operation) -100 … 275 ° F -73 … 135 ° C  
     
    Temperature coefficient of sensitivity ≤ 0.03 %/°F ≤ 0,054 %/°C [2]
     
    Maximum shock (f &)menu; r fraction of a second) 3,000 ° F 1,650 ° C  
     
    maximum acceleration 2,000 g pk 19.600 m/s² PK  
     
    Maximum shock resistance 20,000 g pk 196,000 m/s² pk  
    Technical characteristics
    Output polarity (positive pressure) positive    
     
    Discharging charging time constant (at room temperature) ≥ 1.0 s    
     
    power 22 … 30 V DC    
     
    Constant power supply 2 … 20 mA    
     
    output impedance < 100Ω    
     
    Bias voltage 8 … 14 V DC    
     
    electrical insulation 100 MΩ  
    -physical Properties
    geometry of the sensor element compression principle    
     
    Sensor element quartz    
     
    housing material stainless steel    
     
    membrane INVAR    
     
    Sealing hermetically sealed welded    
     
    electrical connection 10-32-coaxial socket    
     
    Mass 0.50 oz 14.3 grams  
    notes
    [1] for output range up to 10 V, the supply voltage must at least 24 V DC be. At low bias voltage of the sensor, no negative output range is usable up to-10 V.    
     
    [2] typically    
     
    [3] zero-based balancing line, determined using the method of least squares    
     
    [4] details: see PCBs Declaration of conformance PS023    
    equipment
    065A03 sealing ring, 0.435 "outside x 0,377" x 0.030 "high brass 3  
    (M) metric mounting
    Accessories      
     
    065A40 sealing ring, 0.435 "outside x 0,397" x 0.030 "high brass 3  
    (S) stainless steel diaphragm
           