PCB-352A24/NC
- ICP®-miniature vibration sensor, li >
- Mass 0.8 grams.
- Mass-insulated,.
- Measurement range 50 g,
- Sensitivity 100 mV/g,
- Frequency range 1... 8,000 Hz,
- 3-56 female connector on the side.
- Without cable
| Basic data | ||||
| Sensitivity (±10 %) | 10.2 mV/(m/s²) | |||
| Measuring range | ±50 gPK | ±490 m/s²PK | ||
| Frequency range (≤5 %) | 1.0... 8,000 Hz | |||
| Frequency range (±10 %) | 0.8... 10,000 Hz | |||
| Frequency range (± 3 dB) | 0.4... 12,000 Hz | |||
| Resonance frequency | ≥30.000 Hz | |||
| Wide band resolution | 0.0002 g RMS | 0,002 m/s² RMS | [1] | |
| Linearity error | ±1 % | [2] | ||
| Transverse acceleration sensitivity | ≤5 % | |||
| Conditions of use | ||||
| Overload limit (shock) | ±5.000 gPK | ±49.050 m/s²PK | ||
| temperature range (operation) | -65... 250 ° F | -54... 121 ° C | ||
| specifications | ||||
| power supply | 18... 30 V DC | |||
| constant power supply | 2... 20 mA | |||
| output impedance | ≤300Ω | |||
| Bias voltage | 8... 12 V DC | |||
| discharging charging time constant | 0,4 … 1,5 s | |||
| settling time (auf Bias-Spannung ±10 %) | >8 s | |||
| spectral noise (1 Hz) | 80 µg/√Hz | 785 (µm/s²)/√Hz | ||
| Spectral noise (10 Hz) | 15 µg/√Hz | 147 (µm/s²)/√Hz | [1] | |
| spectral noise (100 Hz) | 4 µg/√Hz | 39 (µm/s²)/√Hz | [1] | |
| Spectral noise (1,000 Hz) | 1 µg/√Hz | 9.8 (µm/s²)/√Hz | [1] | |
| Electrical insulation (base) | ≥100 MΩ | |||
| Physical properties | ||||
| Size - height | 0.19 in | 4.8 mm | ||
| Size - length | 0.48 in | 12.2 mm | ||
| Size - width | 0.28 in | 7.1 mm | ||
| Mass | 0.03 oz | 0.8 grams | [1] | |
| Sensor element | Ceramic | |||
| Geometry of the sensor element | Shear principle | |||
| Housing material | Anodized aluminium | |||
| Sealing | Epoxy resin | |||
| Electrical connection | 3-56 coaxial connector | |||
| Connecting position | On the side | |||
| Fixing | Adhesive mounting | |||
| Notes | ||||
| [1] | Typically | |||
| [2] | Zero-AusgleichsgeRade, determined using the method of least squares | |||
| [3] | Details: see PCBs Declaration of conformance PS023 | |||
| Included accessories | ||||
| 039A28 | Transfer tool | -1 | ||
| 080A109 | Mount wax | -1 | ||
| ACS-1 | Frequency response, traceable to NIST (10 Hz up to the upper 5%-Punkt) | -1 |