PCB-(M)352C66

PCB-(M)352C66

  • ICP®-miniature vibration sensor,
  • Measurement range 50 g,
  • Sensitivity 100 mV/g,
  • Frequency range 0.5 … 10,000 Hz,
  • Mass 2 grams,
  • 5-44 female connector above
  •  

    basic data
    sensitivity (±10%)10.2 mV/(m/s²) 
     
    range±50 g pk ±491 m/s 2 pk 
     
    frequency range (±5%)0.5 … 10,000 Hz  
     
    Frequency range (± 10%)0.3 … 12,000 Hz  
     
    frequency range (± 3 dB)0.2 … 20,000 Hz  
     
    Resonance frequency≥35,000 Hz  
     
    phase response (at 21 ° C) (±5 °)2 … 6,000 Hz  
     
    wide band resolution0,00016 g RMS0,0015 m/s² RMS[3]
     
    Linearity≤1% [1]
     
    transverse acceleration sensitivity≤5% [2]
    conditions of use
    Overload limit (shock)±5. 000 g pk±49. 050 m/s² pk 
     
    temperature range (operation)-65 …200 °F-54 … 93 ° C[4]
     
    Strain sensitivity (base area)≤0.005 g/µε≤0.05 (m/s²)/µε[3]
    technical characteristics
    Supply voltage18 … 30 V DC  
     
    constant power supply2 … 20 mA  
     
    Output impedance≤300 Ω  
     
    Bias voltage8 … 12 V DC  
     
    discharging charging time constant0.8 … 2.4 s  
     
    Settling time < br(auf Bias-Spannung ±10 %)< 10 s  
     
    spectral noise (1 Hz)60 µg / √Hz588 (µm / s ²) / √Hz 
     
    Spectral noise (10 Hz)16 µg / √Hz157 (µm / s ²) / √Hz[3]
     
    spectral noise (100 Hz)5 µg / √Hz49 (µm / s ²) / √Hz[3]
     
    spectral noise (1,000 Hz)1.5 µg / √Hz14.7 (µm / s ²) / √Hz[3]
    Physical properties
    Size - height0.67 in17.0 mm 
     
    Size - Allen9/32 in7.1 mm 
     
    mass0.070 oz2.0 grams[3]
     
    sensor elementCeramic  
     
    geometry of the sensor elementshear principle  
     
    housing materialtitanium  
     
    Sealinghermetically sealed welded  
     
    electrical connection5-44 coaxial  
     
    connection positionTop  
     
    mounting screw5-40 male  
     
    Assembly torque8 … 12 in-lb90 … 135 NCM 
    Notes
    [1]zero-based balancing line, determined using the method of least squares  
     
    [2]transverse acceleration sensitivity typically ≤ 3%  
     
    [3]typically  
     
    [4]The data for temperatures above 93 ° C applies only for transducers with HT option.  
     
    [5]details: see PCBs Declaration of conformance PS023  
    equipment
    080A109mounting wax-1 
     
    080A15adhesive mounting-1 
     
    ACS-1Frequency response, traceable to NIST (10 Hz up to the upper 5% point)-1 
    (A) glue Assembly
    Accessories   
     
    080A90glue-1 
    (M) metric mounting
    Mounting threadM3 x 0.50 male  
     
    equipment   
     
    M080A15adhesive mounting-1 
    (HT) high temperature electronics
    frequency range(±5%)5 … 10,000 Hz  
     
    Frequency range (± 10%)3 … 12,000 Hz  
     
    frequency range (± 3 dB)2 … 20,000 Hz  
     
    Wide band resolution of0.0002 g RMS0.002 m/s² RMS 
     
    Temperature range-65 … 250 ° F-54 … 121 ° C 
     
    Discharging charging time constant0.08 … 0.24 s  
     
    spectral noise75 µg / √Hz736 (µm / s ²) / √Hz 
     
    Spectral noise25 µg / √Hz245 (µm / s ²) / √Hz 
    (J) mass insulation
    frequency range (±5%)0.5 … 8,000 Hz  
     
    frequency range (± 10%)0.3 … 10,000 Hz  
     
    frequency range (± 3 dB)0.2 … 16,000 Hz  
     
    Resonance frequency  
     
    electricIsolation  
     
    size - hex3/8 in9.5 mm 
     
    size - height0.73 in18.5 mm 
     
    mass0.1 oz2.8 grams 
    (W) waterproof cable connection
    connection positionabove  