PCB-352A73
| basic data | |||
| sensitivity (±20%) | 5 mV / g | 0.51 mV/(m/s²) | |
| range | ± 1.000 g pk | ±9. 810 m/s² pk | |
| frequency range (±5%) | 2.0 … 10,000 Hz | ||
| Frequency range (± 10%) | 1.5 … 25,000 Hz | ||
| frequency range (± 3 dB) | 0.7 … 40,000 Hz | ||
| Resonance frequency | ≥ 70,000 Hz | ||
| wide band resolution | 0.002 g RMS | 0.02 m/s² RMS | [1] |
| Linearity | ≤ 1% | [2] | |
| transverse acceleration sensitivity | ≤ 5% | ||
| conditions of use | |||
| (Shock) overload limit | ± 10.000 g pk | ±98. 000 m/s² pk | |
| Temperature range (operation) | -65 … 250 ° F | -54 … 121 ° C | |
| Specifications | |||
| power supply | 18 … 30 V DC | ||
| Constant power supply | 2 … 20 mA | ||
| output impedance | ≤200Ω | ||
| Bias voltage | 7 … 11 V DC | ||
| Discharging charging time constant | 0.2 … 1,0 s | ||
| settlingbr(auf Bias-Spannung ±10 %) | < 3 s | ||
| Spectral noise (1 Hz) | 875 µg / √ Hz | 8.580 (µm / s ²) / √ Hz | [1] |
| spectral noise (10 Hz) | 150 µg / √ Hz | 1.470 ( µm / s ²) / √ Hz | [1] |
| spectral noise (100 Hz) | 45 µg / √ Hz | 440 (µm / s ²) / √ Hz | [1] |
| spectral noise ( 1,000 Hz) | 20 µg / √ Hz | 196 (µm / s ²) / √ Hz | [1] |
| spectral noise ?10.000 Hz) | 10 µg / √ Hz | 98 (µm / s ²) / √ Hz | [1] |
| physical properties | |||
| size - height | 0.11 in | 2.8 mm | |
| Size - length | 0.34 in | 8.6 mm | |
| size - width | 0.16 in | 4.1 mm | |
| dimensions (without cable) | 0.01 Oz | 0.3 grams | [1] |
| sensor element | ceramic | ||
| geometry of the sensor element | shear principle | ||
| Housing material | titanium | ||
| sealing | hermetically sealed | ||
| wiring | angelö pins coated cable | ||
| cable exit | side | ||
| cable terminating resistance | 10-32-coaxial connector | ||
| Cable length | 10 ft | 3 m | [3] |
| Attachment | bonding | ||
| notes | |||
| [1] | typically | ||
| [2] | zero-based balancing line, determined using the method of least squares | ||
| [3] | composed of 30 cm cable type 089 (32 AWG, twisted pair) and 2.7 m coaxial cable type 018 | ||
| equipment | |||
| 039A26 | Transfer tool | -1 | |
| 080A109 | mounting wax | -1 | |
| ACS-1 | frequency response, Rü ckführbar on NIST (10 Hz up to the upper 5% point) | -1 |