PCB-(M)622A11
basic data | |||
sensitivity (±5%) | 100 mV/g | 10.2 mV/(m/s²) | [4] |
measuring range | ±50 g | ±490 m / s² | |
Frequency range (±5%) | 35 … 240,000 cpm | 0.58 … 4,000 Hz | [5] |
frequency range (± 10%) | 25 … 300,000 cpm | 0,42 … 5,000 Hz | |
frequency range (± 3 dB) | 12 … 600,000 cpm | 0.2 … 10,000 Hz | |
resonance frequency | 1.200 kcpm | 20,000 Hz | [3] |
wide band resolution | 50 µg | 490 µm / s ² | [3] |
linearity | ≤1% | [6] | |
Transverse acceleration sensitivity | ≤5% | ||
conditions of use | |||
Overload limit (shock) | 5,000 g pk | 49,050 m/s² pk | |
temperature range | -65 … 250 ° F | -54 … 121 ° C | |
Protection | IP68 | ||
technical characteristics | |||
settling time (on bias voltage ± 1%) | ≤5.0 s | ||
discharging charging time constant | ≥0.8 s | ||
power | 18 … 28 V DC | ||
constant power supply | 2 … 20 mA | ||
Output impedance | < 100Ω | ||
Bias voltage | 8 … 12 V DC | ||
Spectral noise (10 Hz) | 4.0 µg / √Hz | 39.2 (µm / s ²) / √Hz | [3] |
spectral noise (100 Hz) | 0.8 µg /. ? Hz | 7.85 (µm / s ²) / √Hz | [3] |
spectral noise (1,000 Hz) | 0.4 µg / √Hz | 3.9 (µm / s ²) / √Hz | [3] |
Electrical protection | RFI/ESD | ||
Electrical insulation | >100 MΩ | ||
physical properties | |||
size - hex | |||
[6] | zero-based balancing line, determined using the method of least squares | ||
[7] | Details: See PCBs Declaration of conformance PS023, PS090 and PS061 | ||
Accessories supplied | |||
081A40 | mounting bolts with 1/4 "-28 thread | -1 | |
ICS-1 | frequency response for uniaxial sensors from 10 Hz up to the upper 5% frequency | -1 | |
(M) metric mounting | |||
equipment | |||
M081A61 | metric mounting bolts with M6 thread | -1 | |
(TO) temperature output | |||
temperature range | 36 … 250 ° F | 2 … 121 ° C | |
transcription factor | 5.56 mV / ° F (zero point at 32 °F) | 10 mV / ° C | |
connections (red wire) | Acceleration output | ||
(black wire) pin assigNment | Earth | ||
Pin assigNment (white wire) | temperature output | ||
pin assigNment (green wire) | ground | ||
(EX) with ATEX approval | |||
Approval for use in potentially explosive atmospheres | EEx ia IIC T4,-54 ° C≤Ta≤121 ° C, II 1 G | ||
(LB) low bias voltage | |||
Bias voltage | 6 … 8 V DC | ||
Power | 12 … 28 V DC | ||
measuring range | ±35 g | ±343 m / | |
(CS) CSA-approved | |||
approval for use in potentially explosive atmospheres | CL I, div I, groups A, B, C, D CL II, div I, groups E, F, G Cl III, div I | ||
Approved for use in potentially explosive atmospheres | ExIA IIC T4, AExia IIC, T4 | ||
approval for use in potentially explosive atmospheres | CL I, div 2, groups A, B , C, D ExnL IIC T4, AExnA IIC T4 |