PCB-422E55
    zoom_out_map

      PCB-422E55

    • Inline charge amplifier,
    • Input range of ±10. 000 pC,
    • Output range ±5 V,
    • Transcription factor: 0.5 mV / pC,
    • Frequency range 5 … 100,000 Hz (depends on the constant current)
    • basic data
      transcription factor (±2. 5%)0.5 mV/pC
      range±8 V
      Lower limiting frequency (-5%)0.5 Hz
      upper cut-off frequency (-5%)100,000 Hz[2]
      linearity error≤ 1.0% FS
      conditions of use
      temperature range (operation)-65 … 250 ° F-54 … 121 ° C
      maximum Shock resistance5,000 g pk49,050 m/s² pk
      maximum acceleration (5)100 g pk981 m/s² pk
      technical
      power supply18 … 28 V DC
      constant power supply2 … 20 mA
      Output range±5, 0 V
      Output polarityinverted
      output impedance100 Ω
      Bias voltage9 … 13 V DC
      Maximum sensor input voltage30 V
      wide band noise (1 … 10,000 Hz)33 µV-90 dB
      Spectral noise (1 Hz)9.8 µV / √ Hz-100 dB[1]
      spectral noise (10 Hz)3 µV / √ Hz-110 dB[1]
      Spectral noise (100 Hz)0.8 µV / √ Hz-122 dB[1]
      spectral noise (1,000 Hz)0.4 µV / √ Hz-128 dB[1]
      Spectral noise (10,000 Hz)0.2 MV / √ Hz-134 dB[1]
      capacity (feedback branch of the amplifier)2,000 pF
      Recovery time after clipping10 µs
      discharging charging time constant> 1 s
      Resistance (feedback branch of the amplifier)15,000 M Ω
      influence of the input capacitance (sensor + cable) on transcription factor-0.0005% /. pF
      physical properties
      housing materialstainless steel
      waterproofingepoxy resin
      Electrical connection (input)10-32-coaxial socket
      electrical connection (output)BNC Plug
      Gr öße - diameter0.52 in13 mm
      size - length3.4 in86 mm
      mass1.15 oz32.7 grams
      Notes
      [1]typically
      [2]the upper cutoff frequency can by the level of constant current and the length of the Cable output used be reduced.
      [3]details: see PCBs Declaration of conformance PS024. CE conformity is only by weight &AmericanCatholic.org; provides, if the housing is grounded.