PCB-HT356A66

PCB-HT356A66

  • Triaxial ICP® high-temperature vibration sensor,
  • Continuous use temperature up to 163 ° C,
  • With built-in low-pass filter,
  • Measuring range 500 g
  • Sensitivity 10 mV / g,
  • Frequency range 2 … 4,000 Hz,
  • 4-pin female connector on the side
  • "
    basic data
    sensitivity (±10%)10 mV / g1.02 mV/(m/s²)
    range±500 g pk ±4. 900 m/s² pk
    frequency range (±5%)2 … 4,000 Hz[2] [3]
    Resonance frequency≥ 35,000 Hz
    integrated filterfirst order
    edge steepness of the filter6 dB / Octave[1]
    electric filter corner frequency16,000 Hz[1]
    wide band resolution0.002 g RMS0.02 m/s² RMS[1]
    Linearity≤ ±1%[4]
    transverse acceleration sensitivity≤ 5%
    conditions of use
    Overload limit (shock)±7. 000 g pk±68. 600 m/s² PK
    Temperature range (operation)-65 … 325 ° F-54 … 163 ° C[7]
    strain sensitivity (base area)0.001 g/µε0.01 (m/s²)/µε[1]
    specifications
    power supply22 … 30 V DC[5]
    Constant power supply2 … 20 mA
    output impedance≤200Ω
    Bias voltage 7 … 15 V DC[6]
    discharging charging time constant0.1 … 1,0 s
    settling time
    ( on bias voltage ±10%)
    < 3 s
    spectral noise (1 Hz)450 µg / √ Hz4.415 (µm / s ²) / √ Hz[1]
    Spectral noise (10 Hz)100 µg / √ Hz981 (µm / s ²) / √ Hz[1]
    spectral noise (100 Hz)25 µg / √ Hz245 (µm / s ²)) / √ Hz[1]
    spectral noise (1,000 Hz)20 µg / √ Hz196 (µm / s ²) / √ Hz[1]
    physical properties
    Sensor elementceramic
    geometry of the sensor elementshear principle
    housing materialtitanium
    Sealinghermetically sealed
    size - height0.55 in14.0 mm
    size e length0,80 in20.3 mm
    size - width0.55 in14.0 mm
    weight0.32 oz9.0 grams[1]
    Wiring1/4 "-28 4-pin
    connection positionside
    mounting thread10-32 female thread
    Assembly torque10 … 20 in-lb113 … 225 NCM
    notes
    [1]typically
    [2]The upper cut-off frequency tolerance: ±500 Hz
    [3]the frequency response of the X-axis is limited because the installation method.
    [4]zero-based balancing line, determined using the method of least squares
    [5]the full measuring range can only when sufficient high power supply be used.
    [6]the TEDS option increases the bias voltage to 1 v
    [7]the data for temperatures above 121 ° C applies only for transducers with HT option.
    [8]details: see PCBs Declaration of conformance PS023
    equipment
    080A109mounting wax-1
    080A12adhesive mounting-1
    081 B 05Mounting bolts (10-32 on 10-32)-1
    ACS-1Tfrequency response for Triaxial sensor, r & menu; ckführbar on NIST, 10 Hz up to the upper 5% frequency-1
    M081B05mounting bolts 10-32 to M6 x 0.75-1
    (T) TEDS memory in accordance with IEEE PP1451. 4
    (TLA) TEDS LMS International - free format
    (TLB) TEDS LMS International - format for the automotive industry
    (TLC) TEDS LMS international - format for the aviation industry
    Additional information
    featuresbuilt-in low-pass filter with low cut-off frequency
  • robust housing
  • 22 other articles of the same category: