PCB-WHT356B11/NC
basic data | |||
sensitivity (±10%) | 10 mV / g | 1.02 mV/(m/s²) | |
range | ±500 g pk | ±4. 905 m/s 2 pk | |
frequency (Y - and Z-axis) < br(≤5 %) | 2 … 10,000 Hz | ||
Frequency range (X-axis) (±5%) | 2 … 7,000 Hz | ||
resonance frequency | ≥ 55,000 Hz | ||
Wide band resolution | 0,004 g RMS | 0.04 m/s² RMS | [2] |
linearity error | ≤ 1% | [1] | |
transverse acceleration sensitivity | ≤ 5% | ||
conditions of use | |||
Overload limit (shock) | ±10. 000 g pk | ±98.100 m/s² pk | |
temperature range (operation) | -65 … 325 ° F | -54 … 163 ° C | |
specifications | |||
power supply | 18 … 30 V DC | ||
Constant power supply | 2 … 20 mA | ||
output impedance | ≤200Ω | ||
Bias voltage | 7 … 11 V DC | ||
Discharging charging time constant | 0.3 … 1,0 s | ||
settlingbr(auf Bias-Spannung ±10 %) | < 3 s | ||
Spectral noise (1 Hz) | 1,000 µg / √ Hz | 9.810 (µm / s ²) / √ Hz | [2] |
spectral noise (10 Hz) | 300 µg / √ Hz | 2.943 ( µm / s ²) / √ Hz | [2] |
spectral noise (100 Hz) | 100 µg / √ Hz | 981 (µm / s ²) / √ Hz | [2] |
spectral noise ( 1,000 Hz) | 50 µg / √ Hz | 490 (µm / s ²) / √ Hz | [2] |
physical properties | |||
sensor element | ceramic | ||
Geometry of the sensor element | Shear principle | ||
Housing material | titanium | ||
sealing | hermetically sealed | ||
size e height | 0.4 in | 10.2 mm | |
size - length | 0.4 in | 10.2 mm | |
size - width | 0.4 in | 10.2 mm | |
Dimensions (without cable) | 0.14 oz | 4 grams | [2] |
electrical connection | ,water - and oil-dense integrated cable | ||
Cable exit | side | ||
cable termination | 1/4 "-28 4-pin socket | ||
cable | 5 ft" | 1.5 m | |
cord type | type 034, 4-wire shielded cable | ||
mounting screw | 5-40 female | ||
Assembly torque | 4 … 5 in-lb | 45 … 56 NCM | |
notes | |||
[1] | zero-based balancing line, determined with the method of the smallestSquares | ||
[2] | typically | ||
[3] | details: see PCBs Declaration of conformance PS023 | ||
Accessories | |||
080A | adhesive mounting | -1 | |
080A109 | mounting wax | -1 | |
081A27 | mounting bolts, 5-40 on 5-40 | -1 | |
081A90 | mounting bolts, 10-32 on 5-40 | -1 | |
ACS-1T | frequency response for Triaxial accelerometer, traceable to NIST, 10 Hz up to the upper 5% frequency | -1 | |
M081A27 | metric mounting bolts, 5-40 M3 x 0.50 | -1 | |
information | |||
features | water - and oil-dense integrated cable |