PCB-3713E112G
| basic data | |||
| sensitivity (±3%) | 1,000 mV / g | 101.9 mV/(m/s²) | [1] |
| measuring range | ±2 g pk | pk ±19, 6 m/s² | |
| Frequency range (± 3 dB) | 0 … 400 Hz | [2] | |
| frequency range (± 3 dB) | 0 …200 Hz | [5] | |
| Phase error | < 2,5 ° | ||
| wide band resolution (0,5 … 100 Hz) | [2] | 0.001 m/s² rms | rms 0.1 mg |
| Linearity error | ≤1% | [3] [4] | |
| transverse acceleration sensitivity | ≤5% | ||
| conditions of use | |||
| Overload limit (shock) | ±2. 000 g pk | ±19. 620 m / pk | |
| Temperature range (operation) | -65 … 250 ° F | -54 … 121 ° C | |
| temperature range (storage) | -65 … 250 ° F | -54 … 121 ° C | |
| Temperature coefficient of sensitivity | ≤5% | [6] | |
| Temperature coefficient of sensitivity the zero point | ±4% FSO | [6] [3] | |
| strain sensitivity (footprint) | 0.001 g/µε | 0.01 (m/s²)/µε | [] 2] |
| magnetic sensitivity | 100 | µg/Gs/9.8 (m/s²) / T | [2] |
| technical specifications | |||
| power supply | 6 … 30 V DC | ||
| Power | ? 30 mA | ||
| output impedance | ? 120 ? | ||
| Spannungsoffset | ±160 mV DC | [7] | |
| spectral noise (1 … 100 Hz) | 10 µg / √Hz | 98.1 (µm / s ²) / √Hz | [2] |
| electrical insulation (base) | >100 MΩ | ||
| physical properties | |||
| housing material | titanium | ||
| Sealing | hermetically sealed | ||
| size - length | 0.85 in | 21.6 mm | |
| Size - width | 0.85 in | 21.6 mm | |
| size - height | 0.85 in | 21.6 mm | |
| weight | 0.8 oz | 22.7 grams | |
| Electrical connector | 9-Pin | ||
| connection | on the side | ||
| mounting thread | 10-32 female | ||
| notes | |||
| [1] | measured at 100 Hz | ||
| [2] | typically | ||
| [3] | FSO = "Full Scale Output": maximum output signal of the measuring range (4 V DC). | ||
| [4] | zero-based balancing line, determined using the method of least squares | ||
| [5] | guaranteed; through supplied with Calibration certificate is | ||
| [6] | -54 … 121 ° C, reference 24 ° C | ||
| [7] | offset at 3 meters cable length (type 037) | ||
| [8] | details: see PCB Declaration of conformance PS027 | ||
| equipment | |||
| 080A12 | adhesive mounting | -1 | |
| 081 B 05 | mounting bolts, 10-32 on 10-32 | -1 | |
| M081B05 | mounting bolts, 10-32 to M6 x 0.75 | -1 | |
| ACS-145 T | calibration of amplitude and phase of 2 Hz up to the upper ±3dB point ?200 ?z) | -1 |