PCB-3713B12200G
- Triaxial MEMS accelerometer,
- Suitable for measuring static accelerations,
- Measuring range 200 g
- Sensitivity 10 mV / g,
- Frequency range 0 … 850 Hz,
- Supply voltage 6 … 30 V DC,
- Integrated cable, length 3 m, ending on stripped and tinned ends
| basic data | |||
| Sensitivity (± 10%) | 10 mV / g | 1.0 mV / (m / s) | [1] |
| measuring range | ± 200 g pk | ± 1962.0 m / s pk | |
| Frequency range (± 5%) | 0 ... 850 Hz | ||
| Frequency range (± 10%) | 0 ... 1250 Hz | ||
| resonant frequency | ?11.000 Hz | ||
| Phase response (100 Hz) | <10 ° | ||
| Broadband resolution | 21.1 mg rms | 0.21 m / s RMS | [2] |
| linearity error | ≤1% | [3] | |
| Lateral acceleration sensitivity | ?3% | ||
| Operating conditions | |||
| Overload limit (shock) | ± 3.000 g pk | ± 29,420 m / s² pk | |
| The temperature range (operation) | -65 ... 250 ° F | -54 ... 121 ° C | |
| Temperature range (storage) | -65 ... 250 ° F | -54 ... 121 ° C | |
| Temperature coefficient of sensitivity | ± 3% | [4] | |
| Temperature coefficient of the zero point | ± 2% FSO | [4] [5] | |
| Strain sensitivity (base surface) | 0.0005 g / µε | 0,005 (m / s²) / µε | [2] |
| magnetic sensitivity | 70 ug / µε | 6.9 (m / s²) / T | [2] |
| Technical characteristics | |||
| supply voltage | 6 ... 30 V DC | ||
| current consumption | ?6 mA | ||
| output impedance | ≤100 Ω | ||
| voltage offset | ± 40 mV DC | [6] | |
| Spectral Noise (1) | 2518 µg/√Hz | 24,693 (µm/s²)/√Hz | [2] |
| Electrical insulation (housing) | > 100 M | ||
| physical properties | |||
| housing material | titanium | ||
| seal | hermetically sealed | ||
| Size - Length | 0.8 in | 20.3 mm | |
| Size - Width | 0.8 in | 20.3 mm | |
| Size - Height | 0.8 in | 20.3 mm | |
| Mass (wired) | 4.2 oz | 119 grams | |
| electrical connection | integrated cable | ||
| cable outlet | Laterally | ||
| cable type | Type 037, shielded 10-wire cable | ||
| cable length | 10 ft | 3 m | |
| mounting thread | 10-32 female thread | ||
| Hints | |||
| [1] | Measured at 100 Hz, 10 g RMS | ||
| [2] | typically, | ||
| [3] | Zero-based line of best fit, determined by the method of least squares | ||
| [4] | -54 ... 121 ° C, Reference 24 ° C | ||
| [5] | FSO = "Full Scale Output": maximum output signal over the measuring area (4 V DC ). | ||
| [6] | Offset at 3 meter cable (type 037) | ||
| [7] | Details: see PCB Declaration of Conformance PS027 | ||
| included supply | |||
| 080A12 | adhesive mounting | -1 | |
| 081B05 | Mounting bolts (10-32 to 10-32) | -1 | |
| ACS 103T | Calibration of the amplitude and phase response of 2 Hz to the upper 5% -frequency for triaxial transducer | -1 | |
| M081B05 | Mounting bolts 10-32 on M6 x 0.75 | -1 |