PCB-3713B1150G
basic data | |||
sensitivity (±5%) | 40 mV / g | 4.1 mV/(m/s²) | [1] |
range | ±50 g pk | ±490, 5 m/s 2 pk | |
frequency range (±5%) | 0 … 1,000 Hz | ||
Frequency range (± 10%) | 0 … 1,300 Hz | ||
resonance frequency | ≥ 5,100 Hz | ||
phase response (100 Hz) | < 10 ° | ||
wide band resolution | 6.0 mg rms | 0.059 m/s² RMS | [2] |
Linearitä avoidable | ≤ 1% | [3] | |
transverse acceleration sensitivity | ≤ 3% | ||
conditions of use | |||
Overload limit (shock) | ±3. 000 g pk | ±29. 420 m/s² pk | |
Temperature range (operation) | -65 … 250 ° F | -54 …121 ° C | |
temperature range (storage) | -65 … 250 ° F | -54 … 121 ° C | |
Temperature coefficient of sensitivity | ≤3% | [4] | |
Temperature coefficient of sensitivity the zero point | ±2% FSO | [4] [5] | |
strain sensitivity (base area) | 0.0001 g/µε | 0.001 (m/s²)/µε | [2] |
magnetic sensitivity | 40 | µg/Gs/3.9 (m/s²) / T | [2] |
technical characteristics | |||
Supply | 6 … 30 V DC | ||
power | &le≤6 mA | ||
Output impedance | ≤ 100 Ω | ||
Spannungsoffset | ± 40 mV DC | [6] | |
spectral noise (1) | 600 µg /. √ Hz | 5,890 (µm / s ²) / √ Hz | [2] |
electrical insulation (housing) | ≥ 100 MΩ | ||
physical properties | |||
Housing material | titanium | ||
sealing | hermetically sealed | ||
size - length | 0.8 in | 20.3 mm | |
Size - width | 0.8 in | 20.3 mm | |
size - height | 0.8 in | 20.3 mm | |
Mass | 0.6 oz | 17.3 grams | |
electrical connector | 9-Pin | ||
connection position | side | ||
Mounting thread | 10-32 female thread | ||
notes | |||
[1] | measured at 100 Hz, 10 g RMS | ||
[2] | typically | ||
[3] | zero-based balancing line, determined using the method of least squares | ||
[4] | -54 … 121 ° C, reference 24 ° C | ||
[5] | FSO = "Full Scale Output": maximum output (4 V measuring range DC). | ||
[6] | offset at 3 meters cable length (type 037) | ||
[7] | details: see PCB Declaration of Conformance PS027 | ||
Equipment | |||
080A12 | adhesive mounting | -1 | |
081 B 05 | Mounting bolts (10-32 on 10-32) | -1 | |
ACS-103T | calibration of amplitude and phase response of 2 Hz up to the upper 5% frequency for Triaxial sensor | -1 | |
M081B05 | Mounting bolts 10-32 to M6 x 0.75 | -1 |