PCB-3711B122G
basic data | |||
sensitivity (±5%) | 1,000 mV / g | 101.9 mV/(m/s²) | |
range | ±2 g pk | ±19, 6 m/s 2 pk | |
frequency range (±5%) | 0 … 250 Hz | ||
Frequency range (± 10%) | 0 … 350 Hz | ||
resonance frequency | ≥ 1,300 Hz | ||
phase response (10 Hz) | < 2.5 ° | ||
Wide band resolution | 0.25 mg rms | 0.0025 m/s² RMS | [3] |
Linearitä avoidable | ≤ 1% | [2] | |
transverse acceleration sensitivity | ≤ 3% | ||
conditions of use | |||
Overload limit (shock) | ±3. 000 g pk | ±29. 420 m/s² pk | |
Temperature range (operation) | -65 … 250 ° F | -54 …121 ° C | |
temperature range (storage) | -65 … 250 ° F | -54 … 121 ° C | |
Temperature coefficient of sensitivity | ≤3% | [1] | |
Temperature coefficient of sensitivity the zero point | ±2% FSO | [1] [4] | |
strain sensitivity (base area) | 0.0001 g/µε | 0.001 (m/s²)/µε | [3] |
magnetic sensitivity | 43.7 | µg/GS/4.29 (m/s²) / T | [3] |
technical characteristics | |||
Supply | 6 … 30 V DC | ||
power | &le≤6 mA | ||
Output impedance | ≤ 100 Ω | ||
Spannungsoffset | -40 … 40 mV DC | [5] | |
spectral noise (1) | 22.9 µ g / √ Hz | 225 (µm / s ²) / √ Hz | [3] |
electrical insulation (housing) | ≥ 100 MΩ | ||
physical properties | |||
Housing material | Titanium | ||
Sealed | sealedclose | ||
size - height | 0.45 in | 11.4 mm | |
size - length | 0.85 in | 21.6 mm | |
Size - width | 0.85 in | 21.6 mm | |
dimensions (with cable) | 2.29 oz | 65.0 grams | |
electrical connection | Integrated cable | ||
AKabelabgang | side | ||
cable termination | tinned ends | ||
cable type | Type 010, 4-wire shielded cable | ||
cable length | 10 ft | 3 m | |
mounting | hole | ||
Notes | |||
[1] | -54 … 121 ° C, reference 24 ° C | ||
[2] | zero-based balancing line, determined using the method of least squares | ||
[3] | typically | ||
[4] | FSO = "Full Scale Output": maximum output signal of the measuring range (4 V )(DC). | ||
[5] | offset at 3 meters cable length (type 010) | ||
[6] | details: see PCB Declaration of Conformance PS027 | ||
Equipment | |||
080A152 | mounting clip | -1 | |
081A113 | mounting screw | -2 | |
ACS-85 | calibration of amplitude and phase | -1 | |
M081A113 | metric mounting screw | -2 |