PCB-301A12
- ICP® calibration reference-Shock Accelerometer
- Measuring range 10,000 g
- Sensitivity 0.5 mV / g
- Frequency range 1 ... 10,000 Hz
- 10-32 connector on the side
specifications | |||
---|---|---|---|
basic data | English units | SI units | |
Sensitivity (± 20%) | 0,5 mV/g | 0,051 mV/(m/s²) | |
measuring range | ± 10,000 g pk | ±98.100 m/s²pk | |
Frequency range (± 5%) | 1 … 10.000 Hz | ||
resonant frequency | ≥30.000 Hz | ||
Broadband resolution | 0,02 gRMS | 0,20 | [1] |
Linearitätsfehler (per 1.000 g (9.810 m/s²)) | ≤0,1 % | ||
Lateral acceleration sensitivity | ≤3 % | ||
Operating conditions | |||
Overload limit | ± 12.500 g pk | ±122.625 m/s²pk | |
temperature range | -65 … 250 °F | -54 … 121 °C | |
Technical characteristics | |||
supply voltage | 18 … 30 VDC | ||
Constant current supply | 2 ... 20 mA | ||
output impedance | <100 Ω | ||
Bias voltage | 8 … 12 VDC | ||
discharging | 0,5 … 2,0 s | [1] | |
Physical properties | |||
Size - height | 1,43 in | 36,3 mm | |
Size - hexagon | 5/8 in | 15,9 mm | |
Mass | 1,5 oz | 42 grams | [1] |
sensor element | quartz | ||
Geometry of the sensor element | Scherprinzip | ||
housing material | 17-4 (stainless steel) | ||
seal | Hermetically sealed | ||
Electrical connection | 10-32 coaxial | ||
port location | Laterally | ||
Mounting thread (mounting on shaker) | 1/4 "-28 male thread | ||
Mounting thread (mounting of the test object) | 1/4 "-28 female thread |
[2]
Hints | |||
---|---|---|---|
[1] | typically, | ||
[2] | Details: siehe PCB Declaration of Conformance PS023 |
included supply | ||
---|---|---|
081A08 | Mounting bolts (10-32 to 1/4 "-28) | (2) |
081B20 | Mounting bolt, with shoulder (1/4 "-28 to 1/4" -28) | (2) |
ACS-1 | Frequency response, traceable to NIST (10 Hz to the upper 5% point) | (1) |
ACS-14 | Hopkinson Bar Calibration for shock transducers | (1) |
M081B20 | Mounting bolt 1/4 "-28 to M6 x 0.75 | (2) |