PCB-394C06
| basic data | |||
| vibration frequency (±1%) | 159.2 Hz | ||
| vibration acceleration amplitude (±3%) | 1.00 g RMS | 9.81 m/s² RMS | [4] |
| Vibration speed amplitude | 0.39 in / sec RMS | /9.81 mm / s RMS | [10] |
| Swing path amplitude | 0.39 mil RMS | 9.81 µm RMS | [10] |
| Transversal vibration | ≤3% | ||
| total harmonic distortion (Load: 0 … 99 Gramm) | ≤2% | ||
| total harmonic distortion (Load: 100 …) 149 grams) | ≤5% | ||
| total harmonic distortion (Load: 150 … 210 Gramm) | ≤9% | ||
| maximum load | 7.4 oz | 210 Gram | [8] |
| vibration duration < br(bis zuder automatischen Abschaltung) | 1 … 2.5 minutes | [2] | |
| Battery life (Load: 2 Gramm) | 320 cycles | [3] | |
| battery run time (Load: 25 Gramm) | 600 cycles | [3] | |
| Battery life (Load: 50 Gramm) | 1,600 cycles | [3] | |
| battery run time (Load: 100 Gramm) | 400 cycles | [3] | |
| Battery run time (Load: 150 Gramm) | 160 cycles | [3] | |
| Battery run time (Load: 210 Gramm) | 80 cycles | [3] | |
| conditions | |||
| temperature range (operation) | 15 … 130 ° F | -10 … 55 ° C | |
| technical characteristics | |||
| time | ≤ 3s | [1] | |
| power supply (standard) | battery power supply | ||
| Number of batteries | 4 | ||
| Battery type | AA | [6] | |
| power supply (alternative) | DC power supply | ||
| power supply: voltage | 10 V DC | ||
| Power supply: power | 2.4 A | [5] | |
| battery life (Load: 2 Gramm) | 8 hours | [3] | |
| Battery life (Load: 25 Gramm) | 15 hours | [3] | |
| battery life (Load: 50 Gramm) | 40 hours | [3] | |
| Battery life (Load: 100 Gramm) | 10 hours | [3] | |
| battery life (Load: 150 Gramm) | 4 hours | [3] | |
| Battery life (Load: 210 Gramm) | 2 hours | [3] | |
| physical properties | |||
| size - diameter | 2.2 in | 56 mm | |
| Size - height | 7.8 in | ||
| Weight (with batteries) | 31 oz | 900 grams | [1] |
| mounting thread | 10-32 female thread | [9] [7] | |
| Mounting torque (max) | 10 in-lb | 112 NCM | [7] |
| notes | |||
| [1] | typically | ||
| [2] | automatic shut-offpreset at the factory; on request can be deactivated (see operating instructions) | ||
| [3] | benchmark, based on automatic switch-off time (varies depending on the type of batteries used) | ||
| [4] | factory setting; switchable to 1 g pk (see operating instructions) | ||
| [5] | maximum | ||
| [6] | the use of alkaline batteries is recommended for maximum battery life. | ||
| [7] | internal thread for mounting the sensor | ||
| [8] | total charge including mounting accessories, cables and cable connector | ||
| [9] | tested sensor should be mounted by hand (without tools). | ||
| [10] | calculated reference values | ||
| [11] | details: see PCB Declaration of conformance PS022 | ||
| optional accessory | |||
| 073A16 | power supply for 394 C 06 | ||
| equipment | |||
| 073A15 | battery holder | -1 | |
| 080A109 | mounting wax | -1 | |
| 080A84 | calibration adapter, internal thread external thread 10-32 5-40 | -1 | |
| 080A85 | Calibration adapter, threaded M3 X 0.5 to 10-32 thread | -1 | |
| 081A08 | mounting bolts, 10-32 to 1/4 "28 | -1 | |
| " 081 B 05 | Mounting bolts, 10-32 on 10-32 | 2 | |
| M081B05 | mounting bolts, 10-32 to M6 x 0.75 | -1 | |
| M081B23 | mounting bolts, 10-32 on M5 x 0.80 | -1 |