| basic data | |||
| sensitivity (±2. 0%) | 100 mV/g | 10.2 mV/(m/s²) | |
| range | ±50 g pk | ±490 m/s 2 pk | |
| frequency range (±5%) | 0.5 … 10,000 Hz | ||
| Frequency range (± 10%) | 0.3 … 14,000 Hz | ||
| frequency range (± 3 dB) | 0.2 … 20,000 Hz | ||
| Resonance frequency | ≥ 35,000 Hz | ||
| wide band resolution | 0.004 g RMS | 0.039 m/s² RMS | [2] |
| Linearity | ≤ 1% | [1] | |
| transverse acceleration sensitivity | ≤ 3% | ||
| conditions of use | |||
| Overload limit | ±5. 000 g pk | ±49. 050 m/s² pk | |
| Temperature range | -65… 250 ° F | -54 … 121 ° C | |
| specifications | |||
| power supply | 23 … 30 V DC | ||
| Constant power supply | 2 … 20 mA | ||
| output impedance | < 100 Ω | ||
| bias voltage | 11 … 17 V DC | ||
| discharging charging time constant | 2.0 … 5.0 s | ||
| Settling time < br(auf Bias-Spannung ±10 %) | < 12.0 s | ||
| spectral noise (10 Hz) | 65 µg / √ Hz | 638 (µm / s ²) / √ Hz | [2] |
| Spectral noise (100 Hz) | 20 µg / √ Hz | 196 (µm / s ²) / √ Hz | [2] |
| spectral noise (1,000 Hz) | 15 µg / √ Hz | 147 (µ m/s²) / √ Hz | [2] |
| physical properties | |||
| size - height | 1 ½ in | 38.1 mm | |
| size - hex | 1 3 / 16 in | 30.2 mm | |
| Dimensions | 6.2 oz | 176 grams | [2] |
| sensor element | quartz | ||
| geometry of the sensor element | Shear principle | ||
| housing material | 316 L (stainless steel) | ||
| sealing | hermetically sealed welded | ||
| Electrical connection | 10-32-coaxial socket | ||
| connection position | side | ||
| mounting thread ( Mounting on Shaker) | 1/4 "-28 female | ||
| mounting thread (Assembly of the DUT) | 1/4"-28 female | ||
| notes | |||
| [1] | zero-based balancing line, determined using the method of least squares | ||
| [2] | typically | ||
| [3] | details: see PCBs Declaration of conformance PS023 | ||
| Accessories | |||
| 081A08 | mounting bolts (10-32 to 1/4 "28) | 2 | |
| 081 B 20 | mounting bolts, with shoulder(1/4 "28 1/4"-28) | 2 | |
| ACS-1 | frequency response, traceable to NIST (10 Hz up to the upper 5% point) | -1 | |
| ACS-4 | Low frequency calibration (phase and amplitude) of 0.5 … 10 Hz | -1 | |
| M081B20 | mounting bolts 1/4 "28 M6 x 0.75 | 2 | |
| (M) metric montage | "|||