TMS-9210D
zoom_out_map

TMS-9210D

  • Portable system for calibration of vibration sensors
  • Frequency range 0.7 ... 2,000 Hz
  • adjustable amplitude
  • LCD display in SI or US units
  • Battery operation up to 18 hours
  • Integrated ICP® supply and display of the bias voltage
  • Signal input socket and output socket for reference signal
  • Including data storage and USB port for data export
Specifications
Basic dataEnglish unitsSI units 
Maximum amplitude - vibration acceleration 2 g pk 305 m/s²pk
Maximum amplitude - vibration speed 12 in / sec pk 305 mm/spk
Maximum amplitude - displacement 200 mils pk-pk 5 mm pk-pk
Maximum sensor mass 800 grams [1]
Frequency range 42 … 120.000 cpm 0.7 ... 2,000 Hz
Measurement uncertainties
Vibration acceleration
(2 ... 2,000 Hz)
± 3% [2] [7]
Vibration acceleration
(0.7 ... 2,000 Hz)
± 10% [2] [7]
Vibration speed
(3 ... 15 Hz)
± 3% [2] [7]
Vibration path
(3 ... 15 Hz)
± 3% [3]
Vibration path
(3 ... 150 Hz)
± 2 dB [3]
Amplitudenlinearität (100g Payload, 100Hz) <1%
Distortion (1… 5Hz) Typically <15%
Distortion (> 5… 20Hz) Typically <10%
Distortion (> 20… 2,000Hz) Typically <7%
Displayable units
Vibration acceleration gpk, gRMS, m/s²pk, m/s²RMS
Vibration velocity mm/spk, in/spk
Oscillation path mmpk-pk, milspk-pk
frequency Hz, CPM
Test sensor sensitivity mV, EU [4]
Conditions of use
Temperature range 32 ... 122 ° F 0 ... 50 ° C
Inputs / outputs
Integrated ICP® supply And
Input for sensor to be tested ICP, voltage
Test of the bias voltage
(for ICP® sensors)
And
Indication of the sensitivity of the test object in the integrated display And
Input for external shaker excitation Ja (1 VAC rms)
Output of the reference sensor signal Yes (100 mV / g)
power supply
Internal battery 12 VDC, 4 Ah
External supply 110… 240 V, 50… 60 Hz
Working time with battery
(100 grams mass, 100 Hz, 1 g pk )
14 hours [6]
Working time with battery
(100 grams mass, 100 Hz, 10 g pk )
7 hours [6]
Physical characteristics
Size - height 8.5 in 22 cm
Size - width 12 in 30.5 cm
Size - depth 10 in 28 cm
Dimensions 18 lb 8.2 kg
Sensor mounting thread 1/4 "-28 internal thread

Hints
[1] Amplitude and frequency range restricted with large sensor mass (see operating instructions)
[2] Measured with a reference sensor with a mass of 30 g
[3] Measured with a laser interferometer
[4] EU can be [g], [m / s2], [mm / s], [mils] or [µm]
[5] 5 mA constant current supply for ICP® and IEPE sensors
[6] In new condition
[7] Depending on the payload at higher frequencies, lateral movement can cause increased measurement uncertainty
TMS-9210D

Specific References