TMS-9210D
    zoom_out_map

      TMS-9210D

      • Portable system for calibration of vibration sensors
      • Frequency range 0.7 ... 2,000 Hz
      • adjustable amplitude
      • LCD display in SI or US units
      • Battery operation up to 18 hours
      • Integrated ICP® supply and display of the bias voltage
      • Signal input socket and output socket for reference signal
      • Including data storage and USB port for data export
      Specifications
      Basic dataEnglish unitsSI units 
      Maximum amplitude - vibration acceleration2 g pk305 m/s²pk
      Maximum amplitude - vibration speed12 in / sec pk305 mm/spk
      Maximum amplitude - displacement200 mils pk-pk5 mm pk-pk
      Maximum sensor mass800 grams[1]
      Frequency range42 … 120.000 cpm0.7 ... 2,000 Hz
      Measurement uncertainties
      Vibration acceleration
      (2 ... 2,000 Hz)
      ± 3%[2] [7]
      Vibration acceleration
      (0.7 ... 2,000 Hz)
      ± 10%[2] [7]
      Vibration speed
      (3 ... 15 Hz)
      ± 3%[2] [7]
      Vibration path
      (3 ... 15 Hz)
      ± 3%[3]
      Vibration path
      (3 ... 150 Hz)
      ± 2 dB[3]
      Amplitudenlinearität (100g Payload, 100Hz)<1%
      Distortion (1… 5Hz)Typically <15%
      Distortion (> 5… 20Hz)Typically <10%
      Distortion (> 20… 2,000Hz)Typically <7%
      Displayable units
      Vibration accelerationgpk, gRMS, m/s²pk, m/s²RMS
      Vibration velocitymm/spk, in/spk
      Oscillation pathmmpk-pk, milspk-pk
      frequencyHz, CPM
      Test sensor sensitivitymV, EU[4]
      Conditions of use
      Temperature range32 ... 122 ° F0 ... 50 ° C
      Inputs / outputs
      Integrated ICP® supplyAnd
      Input for sensor to be testedICP, voltage
      Test of the bias voltage
      (for ICP® sensors)
      And
      Indication of the sensitivity of the test object in the integrated displayAnd
      Input for external shaker excitationJa (1 VAC rms)
      Output of the reference sensor signalYes (100 mV / g)
      power supply
      Internal battery12 VDC, 4 Ah
      External supply110… 240 V, 50… 60 Hz
      Working time with battery
      (100 grams mass, 100 Hz, 1 g pk )
      14 hours[6]
      Working time with battery
      (100 grams mass, 100 Hz, 10 g pk )
      7 hours[6]
      Physical characteristics
      Size - height8.5 in22 cm
      Size - width12 in30.5 cm
      Size - depth10 in28 cm
      Dimensions18 lb8.2 kg
      Sensor mounting thread1/4 "-28 internal thread

      Hints
      [1]Amplitude and frequency range restricted with large sensor mass (see operating instructions)
      [2]Measured with a reference sensor with a mass of 30 g
      [3]Measured with a laser interferometer
      [4]EU can be [g], [m / s2], [mm / s], [mils] or [µm]
      [5]5 mA constant current supply for ICP® and IEPE sensors
      [6]In new condition
      [7]Depending on the payload at higher frequencies, lateral movement can cause increased measurement uncertainty
      TMS-9210D

      Specific References