PCB-350D12

PCB-350D12

  • ICP® shock acceleration sensor,
  • Masseisoliert,
  • With electric and mechanical filter,
  • Measuring range 50,000 g,
  • Sensitivity 0.1 mV / g,
  • Frequency range 4 … 10,000 Hz,
  • Integrated cable, length 3 m,
    ending on 10-32 plug
  • "

    basic data
    sensitivity (±30%)0.1 mV / g0.01 mV/(m/s²) 
     
    range±50. 000 g PK±490. 000 m/s² pk 
     
    Frequency range (± 1 dB)4 … 10,000 Hz  
     
    frequency range (-3 dB)2 … 25,000 Hz [4]
     
    Cutoff frequency of the electrical filter (-3 dB)17,000 Hz [1] [2]
     
    resonance frequency of the mechanical filter45,000 Hz [1] [3]
     
    Resonance frequency≥ 100,000 Hz  
     
    Wide band resolution0.5 g RMS4.9 m/s 2 RMS[1]
     
    Linearity error (per 10,000 g / 98.100 m/s²)≤2.5%  
     
    transverse acceleration sensitivity≤ 7%  
    Conditions of use
    Overload limit (shock)150.000 g PK± 1.471.500 m/s² pk 
     
    Temperature range (operation)0 … 150 ° F-18 … 66 ° C 
     
    temperature range (storage)-40 …200 °F-40 … 93 ° C 
    specifications
    power supply20 … 30 V DC  
     
    Constant power supply2 … 20 mA  
     
    output impedance≤200Ω  
     
    Bias voltage8 … 14 V DC  
     
    Discharging charging time constant0.10 s [1]
     
    settling time < br(auf Bias-Spannung ±10 %)< 1s  
     
    Electrical insulation (housing)> 1 M ?  
    physical properties
    sensor elementceramic  
     
    Geometry of the sensor elementShear principle  
     
    Housing materialtitanium  
     
    sealedsealedclose  
     
    size - hex0.375 in9.5 mm 
     
    size - height1.04 in26.5 mm 
     
    Dimensions (without cable)0.20 oz5.5 grams[1]
     
    electrical connectionintegrated cable  
     
    cable length10 ft3 m 
     
    Cord typetype 032, screened two-core cable with twisted wire  
     
    mounting thread1/4 "-28 thread  
    Instructions
    [1]typically  
     
    [2]second-order low-pass filter  
     
    [3]the Amplitude is increased in the resonance frequency by 9 dB.  
     
    [4]the specified upper limit frequency is a guideline. The frequency response is limited by the combination of electric and mechanical low pass filter.  
     
    [5]details: see PCBs Declaration of conformance PS023. CE conformity is only ensured if the cable shield(green wire) is grounded.  
    Accessories
    ACS-22frequency response, traceable to NIST (100 Hz to ±1dB point)-1 
    (M) metric mounting
    Mounting threadM6 x 0.75 male  

    30 other articles of the same category: