PCB-356B41

PCB-356B41

  • Triaxial vibration sensor seat cushion,
  • For measurements according to DIN EN 30.326,
  • Measuring range 10 g,
  • Sensitivity 100 mV/g,
  • Frequency range 0.5... 1,000 Hz,
  • Integrated cable, length 1.5 m,
    ending on 4-pin female connector.
  • Including extension cable, length 1.5 m,
    ending on three BNC plugs

 

basic data
sensitivity (±10%)100 mV/g10.2 mV/(m/s²) 
 
measuring range±10 g pk±98 m/s 2 pk 
 
Frequency range (±5%)0.5... 1,000 Hz [1]
 
resonance frequency≥27. 000 Hz  
 
wide band resolution0.0002 g RMS0,002 m/s² RMS[3]
 
linearity≤1% [2]
 
transverse acceleration sensitivity≤5%  
conditions of use
Overload limit (shock)±2. 000 g pk±19. 620 m/s² pk 
 
temperature14... 122 ° F10... 50 ° C 
 
temperature effect≤ 0.10 %/°F< 0.17 %/°C[3]
specifications
power supply6.5... 30 V DC  
 
constant power supply0.3... 10 mA  
 
output impedance≤500 Ω  
 
Bias voltage2.8... 4,5 V DC  
 
discharging charging time constant1... 3 s  
 
settling time
(auf Bias-Spannung ±10 %)
>20 s  
 
spectral noise (1 Hz)50 µg/√Hz490 (µm/s²)/√Hz[] 3]
 
spectral noise (10 Hz)10 µg/√Hz98 (µm/s²)/√Hz[3]
 
spectral noise (100 Hz)4 µg/√Hz39 (µm/s²)/√Hz[3]
 
Spectral noise (1,000 Hz)2 µg/√Hz20 (µm/s²)/√Hz[3]
physical properties
sensor elementceramic  
 
Geometry of the sensor elementShear principle  
 
Housing materialTitanium  
 
SealingHermetic  
 
Size - diameter7.87 in 
 
Size - width0.472 in12 mm 
 
Mass9.6 oz272 grams[3]
 
Electrical connectionIntegrated cable  
 
Cable outletOn the side  
 
Cable termination1/4 "-28 4-pin connector  
 
Cable length5 ft1.5 m 
 
Cable typeType 010, 4-conductor shielded cable  
 
Mounting thread5-40 female  
Notes
[1]Standard calibration to 100 Hz with a sensor in the seat cushion. Special calibration about 100 Hz with sensor built in optional.  
 
[2]Zero-balance just, determined using the method of least squares  
 
[3]Typically  
 
[4]Details: see PCBs Declaration of conformance PS023  
Included accessories
010 G 05Cable-1 
 
081A90Mounting bolts, 10-32 on 5-40-1 
 
ACS-17Calibration of amplitude and phase response, traceable to NIST by 0.5... 100 Hz-1 