PCB-J356A44

PCB-J356A44

  • Triaxial ICP® miniature vibration sensor
  • Mass-insulated housing
  • With TEDS 1.0
  • Measuring range 100 g
  • Sensitivity 50mv / g
  • Frequency range 0.4... 10,000 Hz
  • Dimensions 4.2 grams
  • 4-pin female connector on the side

 

Specifications
Basic dataEnglish unitsSI units 
Sensitivity (±10 %)50 mV / g5.1 mV/(m/s²) 
Measuring range±100 g pk±981 m/s² pk 
Frequency range (≤5 %)0.7... 7,000 Hz 
Frequency range (±10 %)0.4... 10,000 Hz 
Resonance frequency≥30.000 Hz 
Wide band resolution (1... 10,000 Hz)0.0006 g rms0.006 m/s² rms[1]
Linearity error≤1 %[2]
Transverse acceleration sensitivity≤6 % 
TEDS memory chips in accordance with
IEEE P1451. 4
Yes 
Conditions of use
Overload limit (shock)±5.000 g pk±49.050 m/s² pk 
Temperature range (operation)-65 … 200 °F-54... 93 ° C 
Strain sensitivity (footprint)0.001 g/µε0.01 (m/s²)/µε[1]
Technical characteristics
Power supply20... 30 V DC 
Constant power supply2... 20 mA 
Output impedance≤200Ω 
Bias voltage8... 12 V DC 
Discharge charge time constant0.8... 2.4 s 
Settling time (on area within ±10% of the bias voltage) 
Spectral noise (1 Hz)1,962 (µm/s²)/√Hz[1]
Spectral noise (10 Hz)45 µg/√Hz441 (µm/s²)/√Hz[1]
Spectral noise (100 Hz)20 µg/√Hz196 (µm/s²)/√Hz[1]
Spectral noise (1,000 Hz)6 µg/√Hz58.9 (µm/s²)/√Hz[1]
Electrical insulation (base)≥100 MΩ 
Physical properties
Sensor elementCeramic 
Geometry of the sensor elementShear principle 
Housing materialTitanium 
SealingHermetic 
Size - height0.40 in10.2 mm 
Size - length0.40 in10.2 mm 
Size - width0.40 in10.2 mm 
Mass0.15 oz4.2 grams[1]
Electrical connection1/4 "-28 4-pin 
Connecting positionOn the side 
FixingAdhesive mounting 
Notes
[1]Typically  
[2]Zero-balance just, determined using the method of least squares  
[3]Details: See PCBs Declaration of conformance PS023  
Included accessories
080A109Mount wax(1)
080A90Superglue(1)
ACS-1TFrequency response for Triaxial accelerometer, traceable to NIST, 10 Hz up to the upper 5% frequency(1)

30 other articles of the same category: