PCB-(M)109C12

PCB-(M)109C12

  • ICP® pressure sensor with ceramic-coated Cone,
  • For ballistic applications
  • Beschleunigungskompensiert,
  • Measuring range 690.000 kPa,
  • Sensitivity 0.01 mV / kPa,
  • Schockgeschützt up to 50,000 g,
  • 10-32 female connector
  •  

    basic data
    measuring range (output range ±7 V)100,000 psi690,000 kPa 
     
    sensitivity (±15%)0.07 mV / psi0.010 mV / kPa 
     
    Maximum pressure125,000 psikPa 862.000 
     
    resolution2.0 psi13.8 kPa 
     
    Resonance frequency≥400,000 Hz  
     
    rise time≤ 2.0 µs  
     
    linearity≤2.0% FS [1]
    Conditions of use
    Acceleration sensitivity≤0.02 psi/g≤0.015 kPa/(m/s²) 
     
    Temperature range (operation)-100 … 275 ° F-73 … 135 ° C 
     
    Temperature coefficient of sensitivity≤0.07 %/°F? 0,126% /. ° C 
     
    maximum shock (for fraction of a second)3,000 ° F1,650 ° C 
     
    maximum acceleration2.000 g PK PK 
     
    maximum shock resistance50,000 g pk490.330 m/s 2 pk 
    Technical features
    output polarity (positive pressure)positive  
     
    discharging charging time constant (at room temperature) ≥ 2,000 s  
     
    Supply20 … 30 V DC  
     
    constant power supply2 … 20 mA  
     
    output impedance≤100 Ω  
     
    Bias voltage8 … 14 V DC   
    Physical properties
    geometry of the sensor elementcompression principle  
     
    sensor elementquartz  
     
    Housing materialC-300  
     
    membraneC-300 [2]
     
    waterproofingepoxy resin  
     
    Electrical connection10-32-coaxial socket  
     
    weight0.6 oz17 grams 
    notes
    [1]zero-based balancing line,. determined with the method of least squares  
     
    [2]Keramikbeschichtet  
     
    [3]details: see PCBs Declaration of conformance PS023  
    Equipment
    065A06sealing ring, 0,318 "outside x 0.250" x 0.010 "high, 316 L (stainless steel)3 
    (M) Metric mounting
        

    4 other articles of the same category: