PCB-357C71

PCB-357C71

  • Vibration sensor with charge output,
  • Continuous use temperature up to 482 ° C,
  • Decay,
  • Measuring range 500 g
  • Sensitivity 10 pC/g,
  • Upper frequency limit 4,000 Hz,
  • 2-pin female connector on the side
  • basic data
    sensitivity (±5%)10 pC/g1.02 pC/(m/s²)
    range± 1.000 g pk ±9. 810 m/s² pk
    frequency range (±5%)4,000 Hz[2]
    Resonance frequency≥ 25,000 Hz
    Linearity≤ 1%[4]
    transverse acceleration sensitivity≤ 5%[3]
    Conditions of use
    Overload limit (shock)±2. 000 g pk±19. 620 m/s² pk
    Temperature range-65 … 900 ° F-54 … 482 ° C
    strain sensitivity (footprint)0.033 g/µε 0.32 (m/s²)/µε[1]
    maximum radiation exposure (integrated neutron flux)10.000.000.000 N / cm 2
    maximum Radiation effect(integrated gamma River)100,000,000 wheel
    Technical
    capacity (from pin to pin)525 pF[1]
    capacity (of pin housing)26 pF[1]
    capacity (asymmetry between the pins)≤ 2 pF
    insulation resistance (from pin to pin at 21 ° C)> 1000 M Ω
    Insulation resistance (from pin on housing at 21 ° C)≥ 100 MΩ[1]
    insulation resistance (from pin to pin at 482 ° C)> 100 k Ω
    output polaritydifferential
    physical properties
    sensor elementceramic
    Sealinghermetically sealed
    size - height1.0 in25.4 mm
    size -. Diameter0.75 in19 mm
    size2.6 oz75 g[1]
    wiring7/16-27 2-pin
    connection positionside
    mounting3 holes
    notes
    [1]typically
    [2]the lower cut-off frequency depends on the used signal conditioning electronics.
    [3]transverse acceleration sensitivity typically ≤ 3%
    [4]zero-based balancing line, determined with the method of least squares
    [5]details: see PCBs Declaration of conformance PS081
    Accessories
    081A99screw-3
    ACS-1frequency response, traceable to NIST (10 Hz up to the top 5 percentage point)-1
    (EX) with ATEX approval
    approval for use in potentially explosive atmospheresex ia IIC T4,-54 ° C ≤ Ta ≤ 135 ° C, II 1 G
    Approval for hazardousAtmospheresex ia IIC T1,-54 ° C ≤ Ta ≤ 450 ° C, II 1 G
    approval for use in potentially explosive atmospheresex nL IIC T4,-54 ° C ≤ Ta ≤ 135 ° C, II 3 G
    Approval for use in potentially explosive atmospheresex nL IIC T1,-54 ° C ≤ Ta ≤ 450 ° C, II 3 G