PCB-3711E1110G
basic data | |||
sensitivity (±5%) | 200 mV / g | 20.4 mV/(m/s²) | |
measuring range | ± 10 g | ±98, 1 m/s² pk | |
Frequency range (± 3 dB) | 0 … 1,000 Hz | [2] | |
frequency range (± 3 dB) | 0 … 500 Hz | [3] | |
Phase response | < 10 ° | ||
wide band resolution (0,5 … 100 Hz) | [2] | 0.002 m/s² rms | rms 0.2 mg |
Linearity error | ≤ 1% | [1] | |
transverse acceleration sensitivity | ≤ 5% | ||
conditions | |||
Overload limit (Shock) | ±5. 000 g pk | ±49. 050 m / pk | |
Temperature range (operation) | -65 … 250 ° F | -54 … 121 ° C | |
temperature range (storage) | -65 … 250 ° F | -54 … 121 ° C | |
Temperature coefficient of sensitivity | ±3% | [4] | |
temperature coefficientzero | ±2% FSO | [4] [5] | |
strain sensitivity (footprint) | 0.0004 g/µε | 0.004 (m/s²)/µε | [2] |
Magnetic sensitivity | 100 | µg/Gs/9.8 (m/s²) / T | [2] |
technical specifications | |||
power supply | 6 … 30 V DC | ||
Power | ≤ 10 mA | ||
output impedance | ≤ 120 Ω | ||
Spannungsoffset | ±80 mV DC | [6] | |
spectral noise (1 … 100 Hz) | 45 µg / √ Hz | 441 (µm / s ²) / √ Hz | [2] |
electrical insulation (base) | ≥ 100 MΩ | ||
physical properties | |||
housing material | titanium | ||
Sealing | hermetically sealed | ||
size - height | 0.45 in | 11.4 mm | |
Gr öße - length | 0.85in | 21.6 mm | |
size - width | 0.85 in | 21.6 mm | |
mass | 0.58 oz | 16.3 grams | |
Electrical connection | 4-Pin | ||
Connection | on the side | ||
mounting | two mounting holes | ||
notes | |||
[1] | zero-based balancing line, determined using the method of least squares | ||
[2] | typically | ||
[3] | guaranteed; evidenced by calibration certificate supplied with | ||
[4] | -54 … 121 ° C, reference 24 ° C | ||
[5] | FSO = "Full Scale Output": maximum output (4 V measuring range ( DC). | ||
[6] | offset at 3 meters cable length (type 010) | ||
[7] | details: see PCB Declaration of Conformance PS027 | ||
IncludedAccessories | |||
080A152 | mounting clip | -1 | |
081A113 | mounting screw with 4-40 thread | 2 | |
M081A113 | Mounting screw m2, 5 thread | 2 | |
ACS-145 | calibration of amplitude and phase of 2 Hz up to the upper ±3dB point (500 Hz) | -1 |