PCB-352M208

PCB-352M208

  • ICP®-miniature vibration sensor for vacuum applications,
  • Low outgassing
  • Measuring range 500 g
  • Sensitivity 10 mV / g,
  • Frequency range 2 … 10,000 Hz,
  • Mass 0.7 grams,
  • Integrated cable, length 3 m,
    ending on 10-32 plug
  • basic data
    sensitivity (±10%)10 mV / g1.02 mV/(m/s²)
    range±500 g pk ±4. 905 m/s 2 pk
    frequency range (±5%)2 … 10,000 Hz
    Frequency range (± 10%)1 … 17,000 Hz
    resonance frequency≥ 65,000 Hz
    wide band resolution0.003 g RMS0.03 m/s² RMS[1]
    Linearity≤ 1%[2]
    transverse acceleration sensitivity≤ 5%
    conditions of use
    (Shock) overload limit± 10.000 g pk ±98.100 m/s² pk
    Temperature range (operation)-65 … 250 ° F-54 … 121 ° C
    Technical characteristics
    Supply voltage18 … 30 V DC
    Constant power supply2 … 20 mA
    output impedance≤200Ω
    Bias voltage 7 … 11 V DC
    discharging charging time constant0.3 … 1,0 s
    settling time
    (on Bias-Spannung ±10%)
    < 3 s
    spectral noise (1 Hz)1,000 µg / √ Hz9.810 (µm / s ²) / √ Hz[1]
    Spectral noise (10 Hz)300 µg / √ Hz2.943 (µm / s ²) / √ Hz[1]
    spectral noise (100 Hz)80 µg / √ Hz785 (µm / s ²) / √ Hz[1]
    spectral noise (1,000 Hz)25 µg / √ Hz308 (µm / s ²) / √ Hz[1]
    physical properties
    size -. Diameter0.24 in6.1 mm
    Größe - height0.32 in8.1 mm
    mass0.03 oz0.7 grams[1]
    sensor elementceramic
    Geometry of the sensor elementshear principle
    housing materialtitanium
    sealing Hermetically sealed
    electrical connectorpins soldered cable
    cable outlettop
    Cable terminating resistance10-32-coaxial connector
    cable length10 ft3 m
    cable typetype 030, coaxial cable
    Attachmentbonding
    notes
    [1]typically
    [2]zero-based balancing line, determined using the method of least squares
    equipment
    080A109mounting wax-1
    080A90Superglue-1
    ACS-1frequency response, traceable to NIST (10 Hz up to the upper 5% point)-1

    4 other articles of the same category: