PCB-(M)602D91
    zoom_out_map

      PCB-(M)602D91

    • Industrial ICP® vibration sensor,
    • With through hole,
    • Measurement range 50 g,
    • Sensitivity 100 mV/g,
    • Frequency range 0.5 … 8,000 Hz,
    • One-point calibration at 100 Hz,
    • M12 connector on the side
    • "
      basic data
      sensitivity (±10%)100 mV/g10.2 mV/(m/s²)[1]
      measuring range±50 g±490 m / s²
      Frequency range (± 3 dB)30 … 480,000 cpm0.5 … 8,000 Hz[4]
      resonance frequency1,500 kcpm25,000 Hz[2]
      Wide band resolution350 µg3.434 µm / s ²[2]
      linearity≤1%[3]
      Transverse acceleration sensitivity≤ 7%
      conditions of use
      (shock) overload limit49,050 m/s ² pkpk 5,000 g
      Temperature range-65 … 250 ° F-54 … 121 ° C
      temperature impact on Temperature coefficient of sensitivitysee curve below[2]
      Protection classIP68
      technical characteristics
      settling time (on area within ±1% of the bias voltage)≤ 3.0 s
      Discharging charging time constant≥ 0.3s
      power18 … 28 V DC
      constant power supply2 … 20 mA
      output impedance< 150 Ω
      Bias voltage8 … 12V DC
      Spectral noise (10 Hz)8.0 µg / √ Hz78.5 (µm / s ²) / √ Hz[2]
      spectral noise (100 Hz)5 µg / √ Hz49,1 (µm / s ²) / √ Hz[2]
      spectral noise (1,000 Hz)4 µg / √ Hz39.2 (µm / s ²) / √ Hz[2]
      Electrical insulation≥ 100 MΩ
      physical properties
      size - length1.55 in39.4 mm
      Size - width0.74 in18.8 mm
      Size - height0,845 in21.5 mm
      mass2.61 oz74.0 grams
      Mounting thread1/4 "-28 threadNo metric equivalent[5]
      Assembly torque2 … 5 ft-lb2.7 … 6.8 Nm
      Sensor elementceramic
      geometry of the sensor elementshear principle
      housing materialStainless steel (316 L)
      Sealinghermetically sealed welded
      connectionon the side
      notes
      [1]conversion factor 1 g = 9.81 m/s²
      [2]typically
      [3]zero-based balancing line,. determined with the method of least squares
      [4]the upper cut-off frequency tolerance: ±10%
      [5]no equivalent in SI
      [6]details: see PCBs Declaration of conformance PS023
      equipment
      081A97 Transit bolt 1 /.4 ' 28-1
      ICS-2single point calibration at 100 Hz (traceable to NIST)-1
      (M) metric mounting
      supplied accessories " ör
      M081A97metric mounting bolt with M6 thread-1
      PCB-(M)602D91

      Specific References

      Download

      602D91_drawing.PDF

      Technical drawing

      Download (97.96k)

      602D91_manual.PDF

      Operating instructions

      Download (927.91k)

      602D91_specs.PDF

      Technical data sheet

      Download (75.26k)