PCB-113B32
    zoom_out_map

      PCB-113B32

    • ICP® high frequency pressure sensor with INVAR-membrane /-housing,
    • Especially insensitive to temperature transients,
    • Beschleunigungskompensiert,
    • Measuring range 34.475 kPa,
    • Sensitivity 0.145 mV / kPa,
    • Lower limiting frequency 1.0 Hz,
    • 10-32 female connector
    •  

      basic data
      measuring range (±5 V output range)5,000 psi34.475 kPa 
       
      usable range (output range ±10 V)10,000 psi68.950 kPa[2]
       
      sensitivity (±0, 05 mV / psi)1.0 mV / psi0.145 mV / kPa 
       
      maximum pressure15,000 psi103.425 kPa 
       
      Resolution0,02 psi0.14 kPa 
       
      resonance frequency≥400,000 Hz  
       
      rise time (reflected pressure)≤ 1.5 µs  
       
      Lower limiting frequency1.0 Hz  
       
      linearity≤1.0% FS [1]
      conditions of use
      Acceleration sensitivity≤ 0.002 psi/g≤ 0,0014 kPa/(m/s²) 
       
      temperature range (operation)-100 … 275 ° F-73 … 135 ° C 
       
      Temperature coefficient of sensitivity≤ 0,1 %/°F≤0.18 %/°C 
       
      maximum shock (for fraction of a second)3,000° F1,650 ° C 
       
      maximum acceleration2,000 g pk19.614 m/s 2 pk 
       
      Maximum shock resistance20,000 g pk196.140 m/s² pk 
      Technical properties
      Output polarity (positive pressure)positive  
       
      Discharging charging time constant (at room temperature)0.3 … 1,0 s  
       
      supply20 … 30 V DC  
       
      Constant power supply2 … 20 mA  
       
      output impedance< 100Ω  
       
      Bias voltage8 … 14 V DC  
      Physical properties
      Geometry of the sensor elementcompression principle  
       
      Sensor elementquartz  
       
      housing materialINVAR  
       
      membraneINVAR  
       
      Sealedsealeddensely welded  
       
      electrical connection10-32-coaxial socket  
       
      mass0.2 oz6.0 grams 
      Notes
      [1]zero-based balancing line, determined using the method of least squares  
       
      [2]for output range supply voltage up to 10 V must be at least 24 V DC. At low bias voltage of the sensor, no negative output range is usable up to-10 V.  
       
      [3]details: see PCBs Declaration of conformance PS023  
      equipment
      060A03installation sleeve, 5 /. 16-24-2A thread, 1/4 "hex, (stainless steel)-1 
       
      060A05installation sleeve, M7 x 0.75-6 g-thread-1 
       
      065A02sealing ring f for flush mounting, 0,248 "outside x 0.219" x 0.015 "high brass3 
       
      065A05gasket spacer, 0,248" outside x 0,221 "x 0.240" high, 17-4 (Stainless steel)-1 
      (S) Emralon coating
      coatingEmralon  
       
      electricIsolation>100 MΩ  
       
      equipment   
       
      065A08insulation ring, 0.250 "outside x 0,218" x 0.027 "high, anodized aluminium3 
       
      065A22seal, 0.250" x outside 0,218 "x 0.015" high, Torlon or Vespel3 
      (W) Watertight cable connection
      equipment   
       
      060A03installation sleeve, 5/16-24-2A thread, 1/4 "hex, (stainless steel)-1 
      (WM) waterproof cable connection
      equipment   
       
      060A05installation sleeve, M7 x 0.75-6 g-thread-1 
      PCB-113B32

      Specific References

      Download

      113B32_drawing.PDF

      Technical drawing

      Download (173.22k)

      113B32_manual.PDF

      Operating instructions

      Download (303.14k)

      113B32_specs.PDF

      Technical data sheet

      Download (31.33k)