PCB-3711B1210G

PCB-3711B1210G

  • MEMS accelerometer,
  • Suitable for measuring static accelerations,
  • Measuring range 10 g,
  • Sensitivity200 mV / g,
  • Frequency range 0 … 1,000 Hz,
  • Supply voltage 6 … 30 V DC,
  • Integrated cable, length 3 m,
    ending on stripped and tinned ends
  • basic data
    sensitivity (±5%)200 mV / g20.4 mV/(m/s²)
    range±10 g pk ±98, 1 m/s 2 pk
    frequency range (±5%)0 … 1,000 Hz
    Frequency range (± 10%)0 … 1500 Hz
    resonance frequency≥ 2.000 Hz
    phase response (100 Hz)< 10 °
    wide band resolution1.2 mg rms0,012 m/s² RMS[3]
    Linearitä avoidable≤ 1%[2]
    transverse acceleration sensitivity≤ 3%
    conditions of use
    Overload limit (shock)±3. 000 g pk±29. 420 m/s² pk
    Temperature range (operation)-65 … 250 ° F-54 …121 ° C
    temperature range (storage)-65 … 250 ° F-54 … 121 ° C
    Temperature coefficient of sensitivity≤3%[1]
    Temperature coefficient of sensitivity the zero point±2% FSO[1] [4]
    strain sensitivity (base area)0.0001 g/µε0.001 (m/s²)/µε[3]
    magnetic sensitivity10.5µg/GS/1.03 (m/s²) / T[3]
    technical characteristics
    Supply6 … 30 V DC
    power&le≤6 mA
    Output impedance≤ 100 Ω
    Spannungsoffset± 40 mV DC[5]
    spectral noise (1)107.9 µg /. √ Hz1,058 (µm / s ²) / √ Hz[3]
    electrical insulation (housing)≥ 100 MΩ
    physical properties
    Housing materialTitanium
    Sealedsealedclose
    size - height0.45 in11.4 mm
    size - length0.85 in21.6 mm
    Size - width0.85 in21.6 mm
    dimensions (with cable)2.29 oz65.0 grams
    electrical connectionIntegrated cable
    Cable exitside
    cable terminationtinned ends
    cable type Type 010, 4-wire shielded cable
    cable length10 ft3 m
    mountinghole
    Notes
    [1]-54 … 121 ° C, reference 24 ° C
    [2]zero-based balancing line, determined using the method of least squares
    [3]typically
    [4] FSO = "Full Scale Output": maximum output signal of the measuring range (4 V )(DC).
    [5]offset at 3 meters cable length (type 010)
    [6]details: see PCB Declaration of Conformance PS027
    Equipment
    080A152mounting clip-1
    081A113mounting screw-2
    ACS-85calibration of amplitude and phase-1
    M081A113metric mounting screw-2

    19 other articles of the same category: