PCB-(M)350A14
- ICP® shock acceleration sensor,
- With an electric filter,
- Measuring range 5,000 g,
- Sensitivity 1 mV / g,
- Frequency range 0.4 … 7,500 Hz,
- 10-32 female connector above
basic data | |||
Sensitivity (± 15%) | 1.0 mV / g | 0.102 mV / (m / s) | |
measuring range | ± 5,000 g pk | ± 49,000 m / s² pk | |
Frequency range (± 10%) | 0.4 ... 7500 Hz | ||
Cut-off frequency of the electric filter (-10%) | ≥7.500 Hz | [3] | |
resonant frequency | ≥50.000 Hz | ||
Broadband resolution | 0.02 g RMS | 0,20 m / s² RMS | [1] |
linearity error | ≤1% | ||
Lateral acceleration sensitivity | ≤5% | [2] | |
Operating conditions | |||
Overload limit (shock) | ± 30,000 g pk | ± 294,000 m / s pk | |
The temperature range (operation) | -65 ... 250 ° F | -54 ... 121 ° C | |
Temperature range (storage) | -65 ... 250 ° F | -54 ... 121 ° C | |
Technical characteristics | |||
supply voltage | 18 ... 30 V DC | ||
Constant current supply | 2 ... 20 mA | ||
output impedance | ≤100 Ω | ||
Bias voltage | 8 ... 12 V DC | ||
discharging | ≥1,8 s | ||
Settling time (on the bias voltage ± 10%) | ?11 s | ||
physical properties | |||
sensor element | quartz | ||
Geometry of the sensor element | shear principle | ||
housing material | stainless steel | ||
seal | hermetically sealed | ||
Size - Hexagon | 0.5 in | 12.7 mm | |
Size - Height | 1.45 in | 36.8 mm | |
Dimensions | 0.63 oz | 17.9 grams | [1] |
electrical connection | 10-32 coaxial | ||
port location | Above | ||
mounting thread | 1/4 "-28 male thread | ||
mounting torque | 25 ... 40 in-lb | 280 ... 450 Ncm | |
Hints | |||
[1] | typically, | ||
[2] | Lateral acceleration sensitivity typically ≤ 3% | ||
[3] | first-order low-pass filter | ||
[4] | Details: see PCB Declaration of Conformance PS023 | ||
included supply | |||
ACS-14 | Hopkinson bar calibration for shock sensor | -1 | |
ACS-22 | Frequency response, traceable to NIST (100 Hz to ± 1 dB point) | -1 | |
(M) Metric assembly | |||
mounting thread | M6 x 0.75 male |