PCB-(M)221B04
    zoom_out_map

      PCB-(M)221B04

    • ICP® force measuring element,
    • Measuring range 4.448 N,
    • Sensitivity 1.12 mV / N,
    • Frequency range 0.001 … 15,000 Hz,
    • 10-32 female connector on the side
    • basic data
      sensitivity (±15%)5 mV / lb1.124,1 mV / kN
      measuring range (pressure)1,000 lb4,448 kN
      Measuring range (traction)1,000 lb4,448 kN
      maximum permissible static load (force)6,000 lb26,69 kN
      maximum static force (traction)1.200 lb5.34 kN
      wide band resolution0,02 lb RMS0,08896 N RMS[1]
      Lower limiting frequency (-5%)0.001 Hz[3]
      upper frequency limit15,000 Hz[4]
      linearity error≤ 1% FS[2]
      conditions
      temperature range-65 … 250 ° F-54 … 121 ° C
      temperature coefficient Temperature coefficient of sensitivity≤ 0.03 %/°F≤ 0,054 %/°C
      technical properties
      discharge charge time constant (at room temperature)&ge≥700 s
      Notes
      [1]typically
      [2]zero-based balancing line, determined using the method of least squares
      [3]calculated on the basis of discharging charging time constant
      [4]suppose value determined on the basis of solid state Dynamics calculations
      [5]details: see PCBs Declaration of conformance PS023
      (M) metric mounting
      (S) negatives Output polarity
      (W) waterproof cable connection
      electrical connectionsealed cable connection
      Cable exitside
      PCB-(M)221B04

      Specific References

      Download

      221B04_drawing.PDF

      Technical drawing

      Download (479.97k)

      221B04_manual.PDF

      Operating instructions

      Download (1.31M)

      221B04_specs.PDF

      Technical data sheet

      Download (69.47k)